User manual
3 Descrizione delle prestazioni
13
3.2.5. Precisione e tempo di risposta
Parametro Precisione
Tempo di
risposta (t
90
)
O
2
±0.2Vol.% < 20s
CO
±20 ppm (0...400ppm)
±5% d. v.m. (401...2000ppm)
±10% d. v.m. (2001...4000ppm)
< 60s
CO, H
2
-comp.
±10ppm o ±10% d. v.m.
2
(0...200ppm)
±20 ppm o ±5% d. v.m.
2
(201...2000ppm)
±10% d. v.m. (2001...8000ppm)
< 40s
COlow, H
2
-comp. ±2 ppm (0…39,9ppm)
±5% d. v.m. (fascia restante)
< 40s
Tiraggio
1
± 0,02 hPa o ± 5% d. v.m.
2
(-0,50...0,60hPa)
± 0,03hPa (0,61...3,00 hPa)
±1,5% d. v.m. (3,01...40,00hPa)
-
Tiraggio
fineFehler!
Textmarke nicht
definiert.
1
± 0,02 hPa o ± 5% d. v.m.
2
(-0,50...0,60hPa)
± 0,03hPa (0,61...3,00 hPa)
±1,5% d. v.m. (3,01...40,00hPa)
∆P (solo con kit
per la pressione
gas 0554 1203)
± 0,5hPa (0,0...50,0hPa)
±1% d. v.m. (50,1...100,0hPa)
±1,5% d. v.m. (fascia restante)
-
Pressione fine
1
(solo con kit per
la pressione gas
0554 1203)
± 0,5hPa (0,0...50,0hPa)
±1% d. v.m. (50,1...100,0hPa)
±1,5% d. v.m. (fascia restante)
Temperatura ± 0,5°C (0,0...100,0°C)
±0,5% d. v.m. (fascia restante)
a seconda
della sonda
Rendimento - -
Perdita calore
sensibile
- -
2
Si applica il valore maggiore