User's Manual

7 Maintaining the product
32
Do not remove the auxiliary circuit board for the NO
sensor until immediately before installation. Do not
leave the sensor without auxiliary circuit board for
longer than 15 minutes.
3. Install new sensor/new bridge in the slot:
Slot 1: O
2
sensor
Slot 2: CO sensor or CO
low
sensor
Slot 3: NO sensor or NO
low
sensor
4. Push the hose connections onto the sensor/bridge.
5. Close the measuring instrument.
When retrofitting a sensor, the associated measurement
parameter/unit must be enabled in the reading display.
Pos: 57 /TD/Produkt ins tand halten/testo 330 und t esto 350 neu/testo 330/ 350 Instandhaltung Ab gassonde reinigen @ 6\ mod_1270039302035_79. docx @ 60667 @ 2 @ 1
7.7. Cleaning the modular flue gas probe
Disconnect the flue gas probe from the measuring instrument
prior to cleaning.
1. Release the probe catch by pressing the key on the probe
handle and remove the probe module.
2. Blow compressed air through the flue gas ducts in probe
module and probe handle (see illustration). Do not use a brush!
3. Fit a new probe module on the probe handle and engage it in
place.
Pos: 58 /TD/Produkt ins tand halten/testo 330/ testo 330 Sondenmodul wechseln @ 6\mod_12792631 31873_79.docx @ 66807 @ 2 @ 1
7.8. Replacing the probe module
Disconnect the flue gas probe from the measuring instrument.