7 Maintaining the product
34
1. Open the filter chamber: turn slightly anti-clockwise (1). Remove
the filter chamber (2).
2. Remove the filter plate (3) and replace it with a new one (4
[0554 3385]).
3. Attach the filter chamber and lock it: turn slightly clockwise.
Pos: 60 /TD/Produkt ins tand halten/testo 330 und t esto 350 neu/testo 330/ 350 Instandhaltung Thermoelement wechseln @ 6\mod_1 270039469873_ 79.docx @ 60729 @ 2 @ 1
7.10. Changing the thermocouple
1. Release the probe catch by pressing the key on the probe
handle and remove the probe module.
2. Remove the thermocouple plug-in head from the socket using a
screwdriver and pull the thermocouple out of the probe shaft.
3. Keep inserting the new thermocouple into the probe shaft until
the connection head clicks into place.
4. Fit a new probe module on the handle and engage in place.
Pos: 61 /TD/Tipps und Hilf e/testo 330i Firm ware-Update @ 18\mod_14383592 71696_79.docx @ 217194 @ 2 @ 1
7.11. Updating the instrument software
At www.testo.com/download-center you can download the current
instrument software (firmware) for the testo 330i.
The measuring instrument must be switched off.
1. Plug the measuring instrument mains unit into a mains socket.
2. Press and hold down for 10 s.
- Both status LEDs (blue/red) slowly flash alternately.
3. Insert the connecting cable (0449 0047) into the USB port on
the measuring instrument, then connect it to the PC.
- Your PC identifies the measuring instrument as a removable
medium.
4. Copy the new instrument software file (ap330ir.bin) to the
identified removable medium.
- Both status LEDs (blue/red) quickly flash alternately. This
process may take a few minutes.
5. Disconnect the connecting cable from the measuring
instrument.
- Once the instrument software has been updated, the measuring
instrument will automatically reboot and is ready for use again.