Datasheet

Setup
2.3.1. 3-Lead ECG Application
The following procedures show a quick method to assemble the ADS1293 and perform an evaluation
for a 3-Lead configuration using an ECG Simulator.
Connect the following electrodes to the ADS1293EVM:
Connect RA to IN1
Connect LA to IN2
Connect LL to IN3
Connect RL to IN4
Figure 7: 3-LEAD ECG Connection
Open the ADS1293 software. On the upper left hand side of the GUI, under the Lead
Configuration field, choose the “3-Lead” application.
After the 3-lead application has been selected, the software will automatically change the
registers for this application. To see what registers have been changed, click on the
Register Map” icon located on the upper left hand of the GUI (see image below).