Datasheet

Using the ADS1298ECG-FE Software
www.ti.com
3.4.4 GPIO and OTHER Registers Tab (ADC Register)
The GPIO and Other Registers tab, located under the Analysis tab, includes controls for GPIO1 through
GPIO4, respiration phase and frequency, routing of the Wilson amplifiers, and derivation of the Goldberger
terminals. Figure 14 shows the GPIO and OTHER Registers tab and all controls contained on the tab.
Figure 14. GPIO and OTHER Register Tab
The General-Purpose I/O Register (GPIO) controls the four general-purpose I/O pins. Each GPIO can be
set as an input or an output via GPIOCx controls. If the output is selected, the GPIODx control is enabled
allowing the user to set the value to output. If the GPIO is selected as an input, the GPIODx control is
disabled and shows the value of the GPIO. If any of the GPIOs are selected as inputs, the Read GPIO
control is enabled which allows the GPIODx values to be updated to the current GPIO value.
The PACE Detect Register does not enable a special PACE measurement mode. The register allows for
enabling and configuration of the PACE amplifiers. PACE Amplifier 1 can connect to input channels 1-4
and Pace Amplifier 2 can connect to input channels 5-8.
The Configuration 4 Register allows control over the Respiration Frequency, WCT connection to the
RLD and lead-off comparator enable status.
NOTE: The Respiration Frequency control is disable since the functionality is not available on the
ADS1298 and ADS1198.
The Respiration Control Register is disabled for the ADS1298 and not available for the ADS1198.
20
ADS1298ECG-FE/ADS1198ECG-FE SBAU171CMay 2010Revised September 2012
Submit Documentation Feedback
Copyright © 2010–2012, Texas Instruments Incorporated