ADS1298ECG-FE/ADS1198ECG-FE Overview
www.ti.com
1.2 Introduction
The ADS1x98ECG-FE is intended for evaluating the ADS1298 and ADS1198 for ECG and EEG
applications. The digital SPI™ control interface is provided by the MMB0 Modular EVM motherboard that
connects to the ADS1x98ECG FE evaluation board. The ADS1x98ECG-FE (see Figure 1) is NOT a
reference design for ECG and EEG applications; rather, its purpose is to expedite evaluation and system
development. The output of the ADS1298 yields a raw, unfiltered ECG signal.
The MMB0 motherboard allows the ADS1x98 to be connected to the computer via an available USB port.
This manual shows how to use the MMB0 as part of the ADS1x98ECG-FE, but does not provide technical
details about the MMB0 itself.
This document covers the operation of the ADS1x98ECG-FE evaluation system. Throughout the
document, the abbreviation EVM and the term evaluation module are synonymous with the
ADS1x98ECG-FE . For clarity of reading, this manual will refer only to the ADS1298ECG-FE or
ADS1x98ECG-FE, but operation of the ADS1198ECG-FE is identical, unless otherwise noted.
CAUTION
Many of the components on the ADS1x98ECG-FE are susceptible to damage
by electrostatic discharge (ESD). Customers are advised to observe proper
ESD handling precautions when unpacking and handling the EVM, including
the use of a grounded wrist strap, bootstraps, or mats at an approved ESD
workstation. An electrostatic smock and safety glasses should also be worn.
1.3 Supported Features
Hardware Features:
Configurable for bipolar or unipolar supply operation
Configurable for internal and external clock and reference via jumper settings
Configurable for AC- or DC-coupled inputs
Configurable for up to 12 standard ECG leads
External Right Leg Drive (RLD) Reference (V
CC
– V
EE
)/2
External Wilson central voltage
Easy connectivity to popular ECG simulators
Software Features:
Designed to display 12 lead ECG data
Analysis tools including a virtual oscilloscope, histogram, FFT, and ECG display
File printing for post-processing of raw ECG data
Sets the ADS1298/ADS1198 register settings via easy-to-use graphic user interface (GUI) software
Post-processing of ECG data using high-pass, low-pass, and 50/60Hz notch filters
1.4 Features Not Supported in Current Version
Real-time data processing
AC lead-off detection filters
QRS detection algorithms
Software PACE detection algorithms
1.5 ADS1x98ECG-FE Hardware
Figure 1 shows the hardware included in the ADS1x98ECG-FE kit. Contact the factory if any component is
missing. The latest software is available for download on the TI website at http://www.ti.com.
8
ADS1298ECG-FE/ADS1198ECG-FE SBAU171CMay 2010Revised September 2012
Submit Documentation Feedback
Copyright © 2010–2012, Texas Instruments Incorporated