User's Manual

BiWaze® Clear System - User Manual Page | 12
I Therapy section displays information about the therapy being delivered.
J Play / Pause icon allows the user to start and pause the therapy.
K
Device Settings icon allows the user to make some configuration changes, view log files and access
administrative screens.
L
Edit icon allows the user to edit Auto Therapy. This icon is only visible for Auto Therapy Mode and if the
software is unlocked or Allow User Adjustment is enabled in Lock Device Options.
5.1 Therapy Modes
Therapy can be delivered in two modes:
Therapy Description
Auto In this mode, the therapy delivery is preprogramed and require no/minimal human intervention
throughout the therapy. There can be up to 10 pre-programmed therapies available in the drop-
down list under Auto. Each therapy is a combination of therapy phases (PEP, OSC and NEB)
and each phase can have a different set pressure and time. Each therapy phase is initiated
after the completion of the previous phase for the set time. This mode is suitable for the
majority of applications whereby users with various levels of training would be able to deliver
the therapy with proficiency.
Manual In this mode, the therapy phase is initiated based on user input through LCD display
touchscreen. The device delivers the set PEP, OSC and NEB therapy based upon the
pressure set for PEP and OSC (NEB pressure is not adjustable) and the duration is based
upon the duration of time set by the user. This mode is suitable to find optimal therapy
parameters for the patient. It is also helpful to use with patients who require frequent pressure
changes during their therapy to assist with comfort and coaching.
Select the therapy mode on the top menu bar.
Note: If the lock icon is blue and closed then the software is locked. If the lock icon is white and open
then the software is unlocked. See Section 7 Locking and Unlocking Feature for more information about
locking and unlocking the software.
5.2 Auto Therapy Mode
Note:
Always use the therapy settings prescribed by the physician.
When BiWaze Clear powers on, the last delivered therapy will be shown on the screen.
If possible, have the patient in an upright position prior to starting therapy.