BiWaze® Clear System - User Manual Page | 16
Note: The total number of phases (F) in
the therapy line and the total therapy
time (G) of all the phases is on the
bottom left.
6.
When the therapy settings are edited to match the prescription, press the green checkmark (H) to save
and close.
7.
The therapy can be saved as a Profile. See Section 6.2 Therapy Profiles for instructions on how to create
a Profile.
5.3 Manual Therapy
Note:
Always use the therapy settings prescribed by the physician.
When BiWaze Clear powers on, the last delivered therapy will be shown on the screen.
If possible, have the patient in an upright position prior to starting therapy.
CAUTION - Ensure the Breathing Circuit and the patient port adapter are dry before delivering the therapy.
5.3.1 Deliver Manual Therapy
1. Ensure the BiWaze Clear control unit and Dual Lumen Breathing Circuit are setup per the instructions in
Section 3 Assemble Dual Lumen Breathing circuit and 4 Power On and Off the Control Unit.
2. Select Manual from the top menu bar. You will notice the LED light around the Play/Pause button will turn
green.
3. Ensure the patient interface is in place on the patient.
Face Maskfirmly, but comfortably, cover the mouth and nose of the patient. Make sure the narrow
end of the mask is over the patient’s nose.
Mouthpieceput the mouthpiece in the patient’s mouth. Make sure that the patient maintains a tight
seal on the mouthpiece during the therapy.
Flexible Tracheostomy Adapterif the patient has a tracheostomy tube or endotracheal tube, use the
flexible tracheostomy adapter to connect the tracheal tube to the system.
4. You can run the therapy on the screen, edit the therapy (if the software is unlocked) or select a therapy
profile from the dropdown list under Manual.
There can be up to 10 therapy profiles. See Section 6.2 Therapy Profiles for instructions.
5. The three therapy phases are each represented by a circle on the screen.
The parameters for PEP and OSC are visible within each therapy circle.
The parameters are editable if the software is unlocked or if Allow User Adjustment is enabled in the
Lock feature.