BiWaze® Clear System - User Manual Page | 35
UI Control Board (UIB): This board
controls the user interface including the
main touch screen LCD as well as USB,
and HDMI ports. This board also provides
wireless interfaces for Wi-Fi connectivity.
Sensor & Nebulizer Board: This board
provides various pressure and flow
sensors required to control as well as
monitor the therapy parameters. This
board also houses the connectors to
peripherals like Nebulizer, USB and HDMI.
Pneumatic block: This block houses the
blowers and valves to deliver air pressure
and flo
w in both positive and negative
direction. The pneumatic paths for
positive and negative flow are
independent.
11.1 Product Specification
OSC Therapy Parameter Modes Min Max
Pressure Manual & Auto 5 cmH
2
0 70 cmH
2
0
Nebulization Manual & Auto Disabled Enabled
Frequency Manual & Auto Low High
Duration Auto 0:05min 30 min
PEP Therapy Parameter Modes Min Max
Pressure Manual & Auto 0 cmH
2
0 30 cmH
2
0
Nebulization Manual & Auto Disabled Enabled
Duration Auto 0:05min 30 min
NEB Therapy Parameter Modes Min Max
Duration Auto 0:05min 30 min
11.2 Environmental
Operating Storage
Temperature 40 F to 95 F (5 C to 35 C) -4° F to 140° F (-25° C to 7 C)
Relative Humidity 15 to 90% (non-condensing) 10 to 90% (non-condensing)
Atmospheric Pressure 106 kPa to 70 kPa 106 kPa to 60 kPa
11.3 Physical
Dimensions
27.5 cm L x 23.5 cm W x 9.0 cm H
(10.5” L x 9.2” W x 3.5” H)
Weight
3.8 kg (8.4 lbs.) (without battery)
4.1 kg (9.4 lbs.) (with battery installed)
11.4 Performance Specifications of the Aerogen® Solo
Flow Rate > 0.2 mL/min (Average ~ 0.38 mL/min)
Particle Size As measured with the Andersen Cascade Impactor:
Specification Range: 1-5 μm
Average Tested: 3.1 μm
As per EN 13544-1:
Aerosol Output rate: 0.30 mL/min
Aerosol Output: 1.02 mL emitted of 2.0 mL dose
Residual Volume: <0.1 mL for 3 mL dose
MCB
UIB
PNEUMATIC
MODULE
Touch
Screen
Display
Sensor Board