User's Manual

BiWaze® Clear System - User Manual Page | 39
function. Therefore, its RF emissions are very low and
are not likely to cause any interference in nearby
electronic equipment.
RF emissions CISPR 11 Class B The device is suitable for use in all establishments,
including domestic establishments and those
directly connected to the public low-voltage power
supply network that supplies building used for
domestic purpose.
Harmonic emissions IEC 61000-3-2 Class A
Voltage fluctuations/Flicker
emissions
IEC 61000-3-3
Complies
12.2 Electromagnetic Immunity
BiWaze Clear is intended for use in the electromagnetic environment specified below. The user of this device
should make sure it is used in such an environment.
Immunity Test
IEC 60601 Test
Level
Compliance Level
Electromagnetic
Environment -
Guidance
Electrostatic Discharge
(ESD)
IEC 61000-4-2
±8 kV contact
±15 kV air
±8 kV contact
±15 kV air
Floors should be wood,
concrete
or ceramic tile. If floors are
covered with synthetic
material, the relative
humidity should be at least
30%.
Electrical fast
Transient/burst
IEC 61000-4-4
±2 kV for power supply
lines
±1 kV for input-output
lines
±2 kV for supply mains
Not Applicable
Mains power quality should
be that of a typical home or
hospital environment.
Surge
IEC 61000-4-5
±1 kV line(s) to line(s)
±2 kV line(s) to line(s)
±1 kV line(s) to line(s)
Not Applicable
Mains power quality should
be that of a typical home or
hospital environment.
Voltage dips, short
interruptions and
voltage variations on
power supply input lines
IEC 61000-4-11
0% UT for 0.5 cycle at:
0
0
, 45
0
, 90
0
135
0
, 180
0
, 225
0
270
0
and 315
0
0% UT for 1 cycle
70% UT for 25/30
cycles, single phase at
0
0
.
0% U
T
for 250/300
cycles
0% UT for 0.5 cycle at:
0
0
, 45
0
, 90
0
135
0
, 180
0
, 225
0
270
0
and 315
0
0% UT for 1 cycle
70% UT for 25/30
cycles, single phase at
0
0
.
0% U
T
for 250/300
cycles
Mains power quality should
be that of a typical home or
hospital
environment. If the user of
the device requires
continued operation during
power mains interruptions,
it is recommended that the
device be powered from an
uninterruptible power supply
or a battery.
Power frequency (50/60
Hz) magnetic field
IEC 61000-4-8
30 A/m 30 A/m Power frequency magnetic
fields should be at levels
characteristic of a typical
home or hospital
environment.
NOTE: UT is the AC mains voltage prior to application of the test level.
Immunity
Test
IEC 60601
Test Level
Compliance
Level
Electromagnetic Environment -
Guidance
Conducted RF
IEC 61000-4-6
3 Vrms
3 Vrms
The BiWaze Clear system is suitable for the
electromagnetic environment of typical homes or
hospital settings.