Datasheet

Buttons
www.ti.com
2 Buttons
There are eight buttons for controlling the output waveform settings. They are listed in Table 1.
Table 1. Buttons
Button Functionality
PHASE_BUTTON Each time the button is pressed, the current INPx pattern becomes the pattern for INNx, the current
INNx pattern becomes the pattern for INPx.
PRF_BUTTON Changes the Pulse Repetition Frequency of the output waveform as given in Table 2 Each press
increases the PRF and then cycles back to 0.
CYCLE_BUTTON Changes the number of times the output pattern cycles within the pulse burst, as given in the Table 3.
Each press moves 1 level up and then back to 0
FREQ_BUTTON Changes the Frequency of the output waveform as given in the Table 4. Each press increases the
frequency and then cycles back to 0
RESET_BUTTON Reset all the setting to its default
PDM_BUTTON One press changes the mode to Power down mode.
CW_MODE_BUTTON First six times the button is pressed will change the pattern depth as shown in Table 5. Seventh press
will change the device to CW mode. In CW mode, pressing the button will revert back to 17-level. The
latch mode is unaffected. Refer to Table 5 and Table 6.
LATCH_MODE_BUTTON First press changes to latch mode. Every consecutive pulse will toggle the mode between Latch and
Non latch mode
Table 2. PRF
Index PRF(kHz) Comments
0 0.5
1 1
2 5 17-level Ramp Only
3 10 17-level Ramp Only
Table 3. Cycle
Index PRF (kHz) Comments
0 1 Longer than 5-level Only
1 2 Longer than 5-level Only
2 3
3 5
4 10
Table 4. Frequency
Index Frequency (MHz) Comments
0 10
1 20
2 50
3 100
10
TX517 Dual Channel, 17-Level With RTZ, Integrated Ultrasound Transmitter SLOU317B August 2011 Revised December 2011
Submit Documentation Feedback
Copyright © 2011, Texas Instruments Incorporated