WIRELESS 433 MHz TEMPERATURE STATION Instruction Manual INTRODUCTION: Congratulations on purchasing this revolutionary 433MHz temperature Station which provides advanced data-logging and data-analysis functions for up to five outdoor temperature readings. The temperature station is connectable to PC by the COM port and data importing to PC is allowed. By applying the smart software provided, user may perform data analysis and statistics conveniently in PC.
• • • • • • • • • • • • • • • • • DCF radio controlled time with manual setting option Time zone option ±12 hours Indoor and outdoor temperature display with MIN/MAX records and time of reception Indoor and outdoor humidity display as RH% with MIN/MAX records Data logging function – with adjustable recording intervals, can store up to 3000 sets of Temperature/Humidity data Connectable to PC (Com port) – data can be transferred to PC with the software (in CD ROM) provided Can receive up to 5 outdoor transmi
The Outdoor Temperature Transmitter (purchased seperately) • • • • Remote transmission of outdoor temperature and humidity to Temperature Station by 433 MHz Shower proof casing Wall mounting case Mounting at a sheltered place. Avoid direct rain and sunshine Stand HOW TO INSTALL AND REPLACE BATTERIES IN THE TEMPERATURE STATION The Temperature Station uses 3 x AA, IEC LR6, 1.5V batteries. To install and replace the batteries, please follow the steps below: 1.
HOW TO INSTALL AND REPLACE BATTERIES IN THE TEMPERATURE TRANSMITTER The Outdoor Temperature transmitter uses 2 x AA. IEC LR6, 1.5V batteries. To install and replace the batteries, please follow the steps below: 1. Remove the battery cover at the front side with an appropriate Battery compartment screwdriver. 2. Insert the batteries, observing the correct polarity (see battery compartment marking). 3. Replace the battery cover on the unit .
BATTERY CHANGE: User shall replace batteries of the temperature station when the battery low indicator is displayed at the LCD. If user does not replace batteries, working error may be resulted. Note: • After changing batteries in temperature station, there is no need for user to reset the temperature station. In fact, the temperature station can remember all transmitter information and sequence as per the pervious set-up.
Note: If the indoor temperature and humidity are not displayed after 15 seconds, remove the batteries and wait for at least 2 minutes before reinserting them. Also check if the battery has been used up. User is recommended to use new alkaline batteries to the units. There is a four-minute time for user to set up all transmitters one by one, after inserting batteries to the temperature station. The below steps 2 to 4 shall be completed within this 4-minute time. 2. 3. 4. 5. 6.
. If after 10 minutes, the DCF time has not been received, press and hold the OK ( ) key for 3 seconds to enter the Set up mode for manual time setting. The clock of the temperature station will automatically attempt to receive the DCF time at midnight. When this is successful, the received time will override the manually set time. The date is also updated with the received time. (Please refer also to notes on “Radio controlled Time Reception” and “Manual Setting”).
FUNCTION KEYS: Temperature Station: There are 5 function keys in the Temperature Station: Plus key OK key Mode key Alarm key OK key ( ) • Minus key In normal display mode, press and hold for 3 seconds to enter the manual setting mode for the following: a. LCD contrast level b. Time zone c. Daylight saving time (On/ Off) d. Radio-controlled time reception (On/ Off) e. 12/ 24 hour Display f. Time setting (Hour and Minute) g. Calendar setting (Year, Month and Day) h.
• i. Recording intervals to memory (hr:min) j. Re-learning/ Removing channels k.
Alarm key (al) • • • Press and hold for 3 seconds to enter the alarm setting mode. In alarm setting mode, press to toggle between the following settings: 1. high humidity alarm 2. low humidity alarm 3. high temperature alarm 4.
LCD Screen: The LCD screen is split into 4 sections displaying the relatively humidity, temperature, EPROM status and Time/ Date. Outdoor data signal reception indicator Relative humidity reading in RH% Outdoor transmitter identification number Temperature reading in ºC/ ºF Data recording interval Memory data display (showing no.
MANUAL SETTINGS: In normal display mode, press and hold the OK key for 3 seconds to enter the manual setting mode for the following: a. LCD contrast setting b. Time zone setting c. Daylight saving time On/ Off setting d. Radio-controlled time reception On/ Off setting e. 12/ 24 hour format setting f. Time setting (Hour and Minute) g. Calendar setting (Year, Month and Day) h. Temperature unit (°C/ °F) setting i. Recording intervals to memory (hr:min) j. Relearning or removing a transmitter k.
LCD CONTRAST SETTING: Digits flashing The LCD contrast can be set within 8 levels, from LCD 0 to LCD7 (Default setting is “LCD 5”): 1. In normal display, press and hold the OK ( ) key until the digits “Lcd” starts flashing. 2. Use the Plus (+) and Minus (–) key to view all levels of contrast. 3. Select the desired LCD contrast. Confirm with the mode or Alarm (al) key to go back to the normal display mode.
TIME ZONE SETTING: Digits flashing The time zone default of the Temperature Station is “0”. To set a different time zone: 1. Press and hold the OK key until the digits “Lcd” starts flashing. 2. Press shortly the OK key one more time to enter the time zone setting. The digits “Zo” will be flashing. 3. Use the Plus and Minus key to set the time zone (-12 to 12 hr). 4. Press the Mode or Alarm key to confirm and go back to the normal display mode.
DAYLIGHT SAVING TIME SETTING Digits flashing 1. 2. 3. 4. Press and hold the OK key until the digits “Lcd” starts flashing. Press shortly the OK key two times to enter the daylight saving time setting. The digit “dS” will be flashing. Use the Plus and Minus key to set the daylight saving time function to “On” or “Off” mode. Press the Mode or Alarm key to confirm and go back to the normal display mode.
TIME RECEPTION ON/OFF SETTING: Digits flashing In area where reception of the DCF time is not possible, the DCF time reception function can be turned OFF. The clock will then work as a normal Quartz clock. (Default setting is ON). 1. Press and hold the OK key until the digits “Lcd” starts flashing. 2. Press and release the OK key three times to enter the time reception time setting. The digit “rEc” will be flashing. 3. Use the Plus and Minus key to set the time reception function to on or off mode. 4.
12/24-HOUR FORMAT SETTING: Digits flashing The time display can be selected to show hours in 12-hour or 24-hour settings. (Default 24-Hour) 1. Press and hold the OK key until the digits “Lcd” starts flashing. 2. Press and release the OK key four times to enter the time format setting. The digit “24h” will be flashing. 3. Use the Plus and Minus key to set the time shown in 12-hour or 24-hour format. 4. Press the Mode or Alarm key to confirm and go back to the normal display mode.
MANUAL TIME SETTING: In case the Temperature Station cannot detect the DCF-signal (for example due to disturbances, transmitting distance, etc.), the time can be manually set. The clock will then work as a normal Quartz clock. Hour flashing Minutes flashing 1. 2. 3. 4. 5. 6. Press and hold the OK key until the digits “Lcd” starts flashing. Press and release the OK key five times to enter the hour setting. The hour digit in the time LCD will be flashing. Use the Plus and Minus key to set the hour.
Note: The unit will still try and receive the signal every day despite it being manually set. When it does receive the signal, it will change the manually set time into the received time. During reception attempts the DCF tower icon will flash. If reception has been unsuccessful, then the DCF tower icon will not appear but reception will still be attempted the following hour. CALENDAR SETTING: Year Date Month The date default of the Temperature Station is 1. 1. in the year 2004.
4. 5. 6. 7. 8. Press the OK key one more time to confirm the year and advance to the month setting. The month digit will be flashing. Use the Plus and Minus key to set the month. Press the OK key one more time to confirm the month and advance to the day setting. The day digit will be lashing. Use the Plus and Minus key to set the day. Press the Mode or Alarm key to confirm and go back to the normal display mode.
RECORDING INTERVAL (hr:min) SETTING Digits flashing The time intervals at which the temperature and humidity data saved to memory can be set to one of the following: 1-minute; 5-minute; 10-minute; 15-minute; 20-minute; 30-minute; 1-hour; 2-hour; 4-hour; 6-hour; 8-hour; 12-hour or 24-hour. Your appliance has been developed and produced to the very latest technical standards. Unfortunately, however, an error occurred in programming the time intervals.
1. 2. 3. 4. Press and hold the OK key until the digits “Lcd” starts flashing. Press and release the OK key eleven times to enter minute setting of the recoding interval. The min in the record LCD will be flashing. Use the Plus and Minus key to set the desired recording interval. Press the Mode or Alarm key to confirm and go back to the normal display mode.
3. 4. 5. 6. 7. Note: Press Plus to get in Learn/ Remove setting of channels. The channel icon “1” will be flashing. Press OK key ( ) to select another channel, if necessary (1-5). Press Plus to set the selected channel to “Re-learn” state (The channel icon will flash in two places on the display), or press Minus to remove the channel (The channel icon will disappear). As soon as the in two places flashing channel number disappears, the reception icon will be flashing in the top left corner.
Up to 5 outdoor transmitters can be received by the temperature station. User may decide the number of transmitter data to be save to the permanent memory of the temperature station. For instant, user may apply 5 transmitters but set to save the data only from particular 3 transmitters. To set the number of transmitters whose data are to be saved in the system memory: 1. Press and hold the OK key until the digits “Lcd” starts flashing. 2.
TOGGLING BETWEEN DIFFERENT CHANNELS (INDOOR AND OUTDOOR CHANNELS): 1. Press the Plus or Minus key to toggle between the Temperature and humidity readings of the indoor channel and the outdoor transmitter channel(s). The channel identification icon (“in”, “out 1”, “out 2”, “out 3” … ) will be displayed at the upper right screen indicating that the particular channel is now being selected. The temperature and humidity readings of Outdoor transmitter No.
VIEWING VARIOUS TYPE OF TEMP/HUMDITY READINGS OF A SELECTED CHANNEL: After a particular channel (the indoor sensor or one of the outdoor transmitters) has been selected by pressing the Plus or Minus key, the following modes of data may be viewed by user: a. The Max temperature and Max humidity Max icon flashing In normal display mode, press the mode key once. The maximum temperature and maximum humidity measured by the channel, and the date and time of the maximum temperature recorded will be displayed.
b. The Min temperature and Min humidity Min icon flashing In normal display mode, press the mode key two times. The minimum temperature and humidity measured by the channel, and the date and time of the minimum temperature recorded will be displayed. The “min” icon will be flashing.
c. The average value of the last 100 sets reading Average icon flashing In normal display mode, press the mode key three times. The average temperature and humidity value of the data which have been saved in the system memory will be displayed. The “avg” icon will be flashing. If the channel is not set to be saved in the system memory, no average value will be estimated and “--.-“ will be shown.
d. The dew point temperature Dew point temperature icon In normal display mode, press the mode key four times. The dew point temperature will be displayed next to the flashing “td” icon.
e. The history data sets of temperature and humidity History data icon flashing In normal display mode, press the mode key five times. The “hist” icon will be flashing and the last recorded temperature and humidity reading with the time of record will be displayed. By pressing the – key, the previous sets of readings which were measured at the user-selected recording intervals may be viewed. If the channel is not set to be saved in the system memory, no history values will be displayed and “--.
CALIBRATION OF TEMPERATURE AND HUMIDITY READING The temperature station allows user to calibrate the displayed temperature and humidity readings by a better reference. User may enter the correction of temperature reading and humidity reading in the range of ±2°C and ±10% respectively. To calibrate the humidity reading: 1. 2. 3. Press the Plus or Minus key to select a channel (the sensor at the temperature station or one of the outdoor transmitters) Press shortly the mode key for six times.
To calibrate the temperature reading: 1. 2. 3. Press the Plus or Minus key to select a channel (the sensor at the temperature station or one of the outdoor transmitters) Press shortly the Mode key for six times. The calibration icon will be flashing. Press shortly the OK key two times. The temperature digits will be flashing. Temperature correction flashing Calibration icon flashing 4. Press the Plus or Minus key to set the desired correction for the humidity reading. Then press the Alarm key to exit.
To set the high humidity alarm: 1. 2. 3. In normal display, press the Plus or Minus key to select a channel (the sensor at the temperature station or one of the outdoor transmitters) Press and hold the Alarm key for about 3 second to enter high humidity alarm setting. The humidity digits, high limit icon and the alarm icon will start flashing. Press the Mode key to select to “activate” or “deactivate” the high humidity alarm.
2. 3. 4. 5. Press and hold the Alarm key for about 3 second to enter high humidity alarm setting. Then press the Alarm key shortly one more time to enter the low humidity alarm setting. The humidity digits, low limit icon and the alarm icon will start flashing. Press the Mode key to select to “activate” or “deactivate” the low humidity alarm. (“Alarm On” or “Alarm Off” icon will be displayed) Press the Plus or Minus key to set the desired value of the low humidity alarm point.
3. 4. 5. Press the Mode key to select to “activate” or “deactivate” the low temperature alarm. (“Alarm On” or “Alarm Off” icon will be displayed) Press the Plus or Minus key to set the desired value of the low temperature alarm point. Press OK key to return to the normal display. Note: • • • • The alarm allow user to set individually the high or low threshold upon his requirements.
Alarm signal icon appears as the current temperature or humidity is out of the pre-set threshold value CLEAR DATA IN THE TEMPERATURE STATION CLEARING THE MAX/MIN TEMPERATURE AND HUMIDITY DATA OF A PARTICULAR CHANNEL To clear the max/min data of a channel: 1. Press the Plus or Minus key to select a channel. 2. Press and hold the Mode key for 3 second. The maximum and minimum temperature/ humidity data and the time of record will be reset to the current value.
1. Press and hold the Mode key for 5 second. The maximum and minimum temperature/ humidity data and the time of record will be reset to the current value. CLEARING ALL STORED TEMPERATURE AND HUMIDITY DATA (FROM TEMPORARY MEMORY OF THE TEMPERATURE STATION) The on-going temperature and humidity data are stored at the ring buffer memory (a temporary memory area where the latest data will displace the oldest data once the memory capacity is reached).
Press and hold OK key first Immediately press these 3 keys at the same time. Then hold the 4 keys together for about 3 seconds. 3. The running digits will be displayed for about 6 minutes, followed by the full segment display. User is then required to remove and re-insert batteries to complete the restart the unit. Note: All data in temperature station will be cleared after the factory reset. The permanent memory of the temperature station will be reset and all the data (of all channels) stored.
The display will show some running digits when the temperature station is reset. Full segment will be displayed- user need to remove and re-insert batteries to restart the temperature station.
After batteries are re-inserted, the station will automatically launch the learn mode. The signal receipt indicator will be displayed. User needs to set again the temperature station for the desired application. (See ”SETTING UP” in previous clause) Note: • • If the battery level is low, the digits “No EPr” may be displayed on LCD after resetting, indicating that the power is too low for the temperature station to function normally. In this case, the batteries shall be replaced.
CHECKING FOR 433MHz RECEPTION If the outdoor temperature data is not being received within three minutes after setting up (or outdoor display show “- -. -” in the outdoor section of the Temperature Station after 3 failed attempts during normal operation), please check the following points: 1. The distance of the Temperature Station or transmitters should be at least 2 meters away from any interfering sources such as computer monitors or TV sets. 2.
POSITIONING THE TEMPERATURE STATION: The Temperature Station provides the option of table standing or wall mounting the unit. Before wall mounting, please check that the outdoor temperature(s) can be received from the desired location(s). To wall mount: 1. Fix a screw (not supplied) into the desired wall, leaving the head extended out by about 5mm. 2. away from the base and hang the station onto the screw. Remember to ensure that it locks into place before releasing.
The mounting surface can affect the transmission range. If e.g. the unit is attached to a piece of metal, it may either reduce or increase the transmitting range. For this reason, we recommend not placing the unit on any metal surfaces or in any position where a large metal or highly polished surface is in the immediate proximity (garage doors, double glazing etc.).
SPECIFICATIONS: Temperature measuring range: Indoor : -9.9ºC to +59.9ºC with 0.1°C resolution +14ºF to +139.8ºF with 0.2ºF resolution (“- -” displayed if outside this range) Outdoor : -29.9ºC to +69.9ºC with 0.1°C resolution -21.8ºF to +157.8ºF with 0.
LIABILITY DISCLAIMER • The electrical and electronic wastes contain hazardous substances. Disposal of electronic waste in wild country and/or in unauthorized grounds strongly damages the environment. • Please contact your local or/and regional authorities to retrieve the addresses of legal dumping grounds with selective collection. • All electronic instruments must from now on be recycled. User shall take an active part in the reuse, recycling and recovery of the electrical and electronic waste.
R&TTE Directive 1999/5/EC Summary of the Declaration of Conformity : We hereby declare that this wireless transmission device does comply with the essential requirements of R&TTE Directive 1999/5/EC.