If the proportional controller type is set to pulse frequency control (PFC):
9. Selecting the relay status under Non-Alarm condition (REL): Press the or
key to choose the desired relay status (de-energised = “DEEN” or energised =
“EN”). Press the ENT key to confirm your selection.
10. Setting the proportional range (PRP): Press the or key to set the
proportional range (setting range: 10 to 500%). Press the ENT key to confirm
your setting.
11. Setting the pulse frequency (F.PF): Press the or key to set the pulse
frequency (setting range: 60 to 120 pulses per minute). Press the ENT key to
confirm your setting.
12. Setting the integral action time (IAT): Press the or key to set the
integral action time (setting range: 0.0 to 999.9 minutes). Press the ENT key to
confirm your setting.
Continue with rest of the Setup procedures by pressing or key or return to the
Measurement mode by pressing the and key (escape) simultaneously.