Instruction Manual αlpha CON 2000
28
6.7 Current Output 2 (CUR 2) sub-function
In this sub-function you set the current output range of the transmitter for temperature values.
SETUP
HOLD
4
SETUP
HOLD
4
SETUP
HOLD
4
SETUP
SETUP
HOLD HOLD
°C
°C
°C °C
44
SETUP SETUP
HOLD HOLD
4
4
ENT ENT ENT
ENT ENT
ENT
ENT
1 2
2
3
3
4
4
1. Select the “CUR 2” sub-function and then press the ENT key.
2. Selecting the output type: press the or key to select the desired output type: 0-20 mA or
4-20 mA. Press the ENT key to confirm your selection.
3. Setting the temperature value at which the transmitter output will be 4 mA (or 0 mA): Press
the or key to set the temperature value (setting range: -10.0 to 115.0 °C or 14 to 237 °F) to
be equivalent to 4 mA (or 0 mA). Press the ENT key to confirm your setting.
4. Setting the temperature value at which the transmitter output will be 20 mA: Press the or
key to set the temperature value (setting range: 0.0 to 125.0 °C or 32 to 257 °F) to be
equivalent to 20 mA. Press the ENT key to confirm your setting.
Continue with rest of the Setup procedures by pressing or key or return to the
Measurement mode by pressing the and key (escape) simultaneously.