29
6.8 Wash relay (WASH) sub-function
In this sub-function you set the parameters for the wash relay.
SETUP
SETUP
HOLD
HOLD
4
4
SETUP SETUP
SETUP
HOLD HOLD HOLD
444
SETUP
SETUP
HOLD
HOLD
4
4
ENT ENT ENT
ENT
ENT ENT
ENT
1 2
2
3
3
4 5
1. Select the “WASH” sub-function and then press the ENT key.
2. Enabling/disabling wash function: Press the or key to enable (WASH ON) or disable
(WASH OFF) wash function. Press the ENT key to confirm your selection.
If “WASH OFF” was selected, the transmitter returns “WASH” sub-function.
If “WASH ON was selected:
3. Selecting the relay status condition: Press the or key to choose desired relay
status (de-energised = “DEEN” or energised = “EN”). Press the ENT key to confirm your
selection.
4. Setting the wash interval in hours: Press the or key to set the desired wash
interval (setting range: 0.1 to 200.0 hours). Press the ENT key to confirm your setting.
5. Setting the wash duration in seconds: Press the or key to set the desired wash
duration (setting range: 1 to 2000 seconds). Press the ENT key to confirm your setting.
Note: During wash cycle the transmitter is set to HOLD. For safety reasons, the
HOLD function is activated 5 seconds prior and deactivated 10 seconds after the
wash cycle.
Continue with rest of the Setup procedures by pressing or key or return to the
Measurement mode by pressing the and key (escape) simultaneously.