User guide
Instruction Manual Alpha DO 500
26
4
As the lower display value changes, the Transmitter adjusts the
upper display reading automatically to suit the new offset value. Up
to ± 10 ºCF offset is allowed.
Press ENT key to confirm the value. The Transmitter reverts to
SET ºC screen.
If ATC disabled (AtC OFF):
55
Setting process temperature: The lower display shows ‘PºC’ and
the upper display shows the last configured process temperature.
Press or key to adjust the upper display to desired process
temperature. Allowable range: 0.0 to 100.0°C.
Press ENT key to confirm the process temperature.
6
Setting calibration temperature: The lower display shows ‘CºC’
and the upper display shows the last configured calibration
temperature. Press or key to adjust the upper display to
desired calibration temperature. Allowable range: 0.0 to 100.0°C.
Press ENT key to confirm the calibration temperature. The
Transmitter reverts to SET ºC screen.
Press or key to access other setup screens or press and key
simultaneously (escape) to return to measurement mode.