User guide
Instruction Manual Alpha DO 500
30
5.8 Viewing Calibration Point
Calibration setting lets you view the high-point at which the calibration has been
performed in the Transmitter.
ENT
SETUP
HOLD
1
ENT
SETUP
HOLD
2
%
11
From measurement mode press ENT key to enter setup mode as
described in section 5.1. The LCD shows the first screen of setup mode
(COF). Press or key to select viewing calibration settings screen
(CAL).
Press ENT key to view calibration settings (CAL).
2
The buffer annunciator
appears in the LCD. The upper display shows
the value of the high-point at which the calibration has been carried out.
The corresponding unit of measurement of calibration is displayed.
NOTE:
If the Transmitter has not been calibrated yet for the currently
selected unit of measurement, the LCD shows an empty
reading (“---“ ) in the upper display.
Press ENT key to exit calibration settings screen. The Transmitter
reverts to SET CAL screen.
Press or key to access other setup screens or press and key
simultaneously (escape) to return to measurement mode.