9.9 Controller Reset – P6.0
This parameter lets you reset the controller. All pH and ORP
calibration data will be cleared, and all Setup parameters revert to
factory default values. Temperature mode (ATC or MTC) and
calibration data remains unchanged.
P6. 0
RsT n0
y
e
S
P6. 1
Press and keys together (ESCAPE) at anytime, to leave the
Setup mode and return to Measurement mode.
P6.0: Controller reset
1. Key in the password “022” as per procedure in Section 5.
2. Press key five times. The screen will scroll P6.0 and RST.
Press ENTER.
3. The screen will scroll P6.1, then NO.
4. Press or keys to toggle between YES and NO. Pressing
YES will clear all calibration data and setup parameters!
5. Press ENTER to confirm selection and return to general Setup
mode.
Press and keys together (ESCAPE) to return to
measurement mode.