- 8 -
2.3.4 Connecting Temperature Probe
For Automatic Temperature Compensated (ATC) pH readings, a 100Ω Pt RTD
temperature probe (2-wire or 3-wire) can be connected to the Controller.
3-Wire Probe:
1. Connect PT100 compensate wire to Pin 5 of J204 connector
2. Connect PT100 sense wire to Pin 6 of J204 connector
3. Connect PT100 GND wire to Pin 7 of J204 connector
2-Wire Probe:
1. Short Pin 5 & 6 of J204 connector using a small piece of wire
2. Connect PT100 sense wire to Pin 6 of J204 connector
3. Connect PT100 GND wire to Pin 7 of J204 connector
2.3.5 Connect Potential Matching Pin (PMP)
If using an electrode with a PMP, connect the additional wire (PMP) from the pH
electrode to Pin 8 of J204 connector. This liquid electrical reference wire is only
used when the pH Controller is configured to ‘Symmetrical mode of operation’.