Integrated Instruction Manual WP 600 Series Meters pH 600 / 610 / 620, COND 600 / 610, DO 600, PC 650, PD 650, CD 650, PCD 650 68X415307 Technology Made Easy ... Part of Thermo Fisher Scientific Rev.
*IMPORTANT – PLEASE READ BEFORE YOU CONTINUE* PLEASE USE THIS MANUAL IN THIS ORDER : Step 1: Use the first half of this manual “General Guide” and “Temperature” to set up and calibrate your instrument (Pages 1 - 39). Step 2: Then go to the specific parameter guides (eg. pH, Conductivity, TDS etc) which are relevant to the model you have purchased. (Pg 41 onwards).
Table of Contents GENERAL GUIDE .................................................................................. 1 1. Overview For All Meters ........................................................................................3 1.1 About the Meters ............................................................................................................................... 3 1.1.1 Display Overview.........................................................................................................
TEMPERATURE ................................................................................... 35 1. Temperature Setup ..............................................................................................37 2. Temperature Calibration ......................................................................................37 2.1 About Temperature Calibration ....................................................................................................... 37 2.
2.9 Calibration Report ........................................................................................................................... 71 2.9.1 To View Calibration Report: ................................................................................................. 71 3. Conductivity Measurement Mode .......................................................................71 3.1 Indicators in Conductivity measurement screen .....................................................................
2.2.1 Auto Salinity Compensation ............................................................................................... 111 3. Concentration in Measurement Mode ..............................................................111 3.1 Indicators in concentration measurement screen ......................................................................... 111 3.2 Dissolved Oxygen Probe............................................................................................................... 112 3.2.
GENERAL GUIDE This section is applicable to all models of the WP 600 Series Meters. Please do not skip this section. Model Parameters pH 600 Temperature (with ATC), pH (-2.00 to 20.00 measuring range). pH 610 Temperature (with ATC), pH (expandable resolution to 0.001). pH 620 Temperature (with ATC), pH (expandable resolution to 0.001) and Ion. COND 600 Temperature (with ATC), Conductivity, TDS. COND 610 Temperature (with ATC), Conductivity, TDS, Salinity, Resistivity.
2
1. Overview For All Meters 1.
1.2 Inserting Batteries NOTE: Please ensure that the gasket is in place otherwise the instrument will not be waterproof. Power up your meter using either: 1. Four ‘AA’ size 1.5 V alkaline batteries (supplied) or, 2. 9V DC power adapter (Optional in some models). 1.2.1 Inserting batteries for the first time 1. Use a Phillips screw driver to remove the four screws holding the battery cover. 2. Insert the batteries in the right direction. 3. Replace the battery cover and screws.
1.3 Attaching the Belt To attach the safety belt: 1. Use a Phillips screw driver to remove the four screws holding the battery cover. 2. Insert the safety belt through the two slots as indicated below. 3. Screw the battery cover back on. Note the ▲UP symbol marked on the cover. 4. Insert your palm between the belt and the body of the meter and adjust the hook & loop fastener.
1.4 Connecting Peripherals 1.4.1 Probes (pH/Conductivity/DO) Attach the probes with correct type of connectors as indicated. pH Probe (BNC) DO Probe with built in temperature sensor (6-pin Connector) Conductivity Probe with built in temperature sensor (8-pin Connector) Use this socket for the standalone temperature probe in the pH only models. Use the electrode properly for best results: 1. Keep the protective plastic electrode guard in tact during measurement and calibration. DO NOT REMOVE IT. 2.
1.4.2 Protective Rubber Boot The rubber boot protects the meter when used in the field. For bench top applications, lift up the stand at the back of the rubber boot. PLEASE SET UP THE SYSTEM BEFORE YOU BEGIN USING THE METER.
2. System Setup Mode For All Meters 2.1 About Setup Mode The setup mode lets you configure various parameters & settings of the meter. You can choose to password-protect your settings, so that other users who may use the meter will not be able to change the settings.
1. Press ENTER key to select Setup Selection screen. 2. Press up or down arrow key to go to required setup sub-group. 3. Press ENTER key to select the currently shown sub-group. Figure 1: Setup Selection Screen Function keys available in setup selection screen: (F1) (Not functional) (F2) (Not functional) (F3) (Not functional) Goes to required setup sub-groups ENTER Selects the current sub-group ESC (F4) Goes to measurement mode from where you entered setup (Not functional) 2.
System setup sub-group allows you to configure general settings of the meter. The settings are displayed in 6 pages. Press NEXT-P (F2) and PREV-P (F1) to navigate through these pages. Parameter Description Factory Default STABLE Indicator ENABLE - The meter displays ‘Stable’ indicator in the measurement screen as per the ‘STABLE CRITERIA’ defined below. DISABLE – ‘Stable’ indicator does not appear.
System Settings Page 2 – Date & Time Figure 3 : System Settings Page 2 – Date & Time This page allows you to set the date & time of the meter.
DISABLE – Meter does not turn off automatically. ON Time After the last key is pressed, no. of minutes the meter should wait before automatically shuts down the meter. Maximum range: 30 min (This parameter has not editable if ‘AUTO OFF’ parameter is disabled) Back Light (permanently ON) ENABLE – Sets the back light always on. DISABLE – Sets the backlight always off. DISABLE Back Light ON with (Key press) ENABLE – The back light of the LCD is automatically on when any key is pressed.
TEXT – Select this format if you use any other method (such as Windows® Hyperterminal) This parameter is used when downloading data from the meter through IrDA Current Data Set TIMED TIMED – Prints measurement data continuously at the interval specified in ‘INTERVAL’ parameter below. SINGLE – Prints only the currently measured reading This parameter applies when PRIN key is pressed from measurement mode to send the currently measured readings to the computer.
a 5-digit password in the ‘SET PASSWORD’ parameter below DISABLE – Disable password protection of the meter Set Pass Word Specify your 5-digit password here. Use (Up) & (Down) key to select a number and then press ENTER key to confirm and move to the next digit. Do not set your password to ‘00000’ as this is reserved for ‘read-only’ password.
Parameter Description Factory Default Clear logged Data Memory YES – Select this to clear all the stored data from the meter’s memory NO – Select this if you do not wish to clear the stored data from the meter’s memory NO Display Contrast Adjust display contrast from 1 - 25 12 Factory Reset Press ENTER if you wish to reset the meter to its factory default settings.
2. The next digit is selected. Press up & down second digit of the password. Enter all 5-digits. arrow keys to enter the 3. Press ENTER key to confirm the password. Note: If you enter an incorrect password, the screen shows “Try Again”. If an incorrect password is entered three consecutive times, the meter goes into measurement mode. If you forget the password, there is no way to access the system setting and calibration. Please contact your authorised dealer for assistance. 1.
3. Calibration Mode For All Meters 3.1 About Calibration The 600 series meters are factory calibrated and allows you to measure pH/mV/ion/conductivity/resistivity/TDS/salinity/DO(%)/DO(mg/L) respective to the model(s) you have purchased. Calibrate to all measurement ranges to ensure the highest accuracy in any given measurement range.
It is recommended that you perform at least a 2-Point Calibration using standard buffers that adequately cover the expected measurement range, prior to measurement. For Ion Connect the ISE to the BNC connector of the meter. Remove plastic protective cap of ISE. Briefly rinse the electrode with clean water to remove any residues. Rinse ISE before and after each calibration or sample measurement to avoid cross-contamination. Ensure that you use new or fresh standard solutions during calibration.
All new calibration values will automatically override the existing data. It is recommended to calibrate the meter periodically and or if it is suspected to be inaccurate. Always rinse the probe with either DI water or rinse solution before and after each calibration/sample measurement. When calibrating in air, make sure that any water droplets from the probe’s membrane are removed. 3.4 Accessing Calibration mode From measurement mode, press CAL (F2) key.
4. Measurement Mode For All Meters 4.
MODE (F3) OFF (F4) Switches between measurement modes Power off the meter (press and hold this key for 3 seconds) Switches between functions groups available in measurement mode ENTER (Not functional) (Not functional) Function Keys available in measurement screen (2nd Group): SETP (F1) MEM (F2) STOR (F3) ESC (F4) Goes to setup mode Shows stored data in the memory Stores the currently displayed reading in the memory Shows 1st Group of functions Function Keys available in measurement screen (3rd Group)
4.2.3 Stable reading indicator You can configure the meter so that LCD displays a ‘Stable’ indicator when the reading does not vary for 2 consecutive seconds. The amount of variations allowed can be set at ‘Slow’, ‘Medium’ or ‘Fast’. 4.2.4 Holding a reading In some situations, you may want to freeze (hold) the measured reading in the LCD for a delayed observation. You can hold a reading in two different ways: Manual Hold – Allows you to hold the reading by pressing HOLD (F1) key at any time you want.
4.4 Alarm set point (For pH/conductivity/DO) You can set the meter to display an alarm when the pH/ conductivity/ salinity/ resistivity/ TDS/ DO (%)/ DO (mg/L) reading goes higher or lower than predefined set points. Refer to ‘Setup’ pages of respective modes to set alarm points for pH, conductivity and DO. Figure 12 : Hi Alarm 4.5 Calibration Due (CAL-DUE) Indicator (For pH/conductivity/DO) You can set a reminder to be displayed in the LCD when the next calibration is due.
5.2.1 Logging data automatically in the meter’s memory 1. Make sure that the Print mode is set to ‘data logging mode’ in the System Setup. You can also specify the time interval for each transfer. 2. From measurement mode, press PRIN (F3) to start saving data automatically in meter’s memory. The memory location of the transferred reading is shown in the bottom-left of the screen. NOTE: This symbol will appear if data logging mode has been selected in “System set up’.
5.2.4 Transferring stored data to the Computer (CyberComm) through IrDA 1. Make sure the CyberComm 600 application is up and running (Refer to the manual in the software packaging to install). 2. Make sure your meter’s IrDA port is facing the IrDA port of the computer. 3. Go to the ‘stored data viewing’ screen as shown above diagram (Use MEM [F2] key. 4. Press PRIN (F2). The screen appears for you to select printing options. or down arrow key to select either ‘all (memory) locations’ or 5.
Transfer Stored data to CyberComm 5.2.5 Transferring stored data to a PC not equipped with infrared receiver. If your PC doesn’t come with a built-in infrared receiver, you can use an RS 232C Adapter (sold separately) that connects directly to your PC using a specially-designed 9-pin plug. Then use the Cybercomm or HyperTerminal software to print your data. You can also use the RS232C Adaptor to print your data directly to our dot matrix microprinter (sold separately) using a custom-built 25-pin plug.
5.2.6 Transferring stored data to a PC using an USB/ irDA Dongle. You can purchase an USB/ irDA dongle 01X447602 (see accessories page at the back of this manual) and transfer data from your meter using the USB port. Please note: This accessory is specially engineered to match your meter’s software to the application in your laptop or PC. If you purchase this item from a different source, it will NOT be compatible for transferring the data.
6. Technical Specifications Temperature (pH and Conductivity) Range Resolution Relative Accuracy Temp. Input Connector -10.0 ºC to 110.0 ºC (14.0 ºF to 230.0 ºF) 0.1 ºC/ 0.1 ºF ± 0.5 ºC / ± 0.9 ºF 8-pin Round Temperature (DO) Range Resolution Relative Accuracy -10.0 ºC to 60.0 ºC (14.0 ºF to 140.0 ºF) 0.1 ºC/ 0.1 ºF ± 0.5 ºC / ± 0.9 ºF Model pH pH 600 Range Resolution Relative Accuracy No. of Calibration points No.
Conductivity Range Measuring ranges Resolution Sub range Accuracy Cal Standards 1 0.050uS to 2.000uS 0.01uS* 1% of FS No 2 2.000uS to 9.990 uS 10.00uS to 99.99uS 100.0uS to 300.0uS 0.01uS* 0.01uS 0.1uS 1% of FS 1% of FS 1% of FS 84.00uS 3 300.0uS to 999.9uS 1.000mS to 4.000mS 0.1uS 0.001mS 1% of FS 1% of FS 1.413mS 4 4.000mS to 9.999mS 10.00mS to 40.00mS 0.001mS 0.01mS 1% of FS 1% of FS 12.88mS 40.00mS to 99.99mS 100.0mS to 500.0mS 0.01mS 0.1mS 1% of FS 1% of FS 111.
Resistivity Range Measuring ranges Resolution Sub range Accuracy 5 10.00MΩ to 20.00MΩ 1.000MΩ to 9.999MΩ 500.0KΩ to 999.9KΩ 0.01MΩ 0.001MΩ 0.1KΩ 1% of FS 1% of FS 1% of FS 4 100.0KΩ to 500.0KΩ 10.00KΩ to 99.99KΩ 3.333KΩ to 9.999KΩ 0.1Ω 0.01Ω 0.001Ω 1% of FS 1% of FS 1% of FS 3 1.000KΩ to 3.333KΩ 250.0Ω to 999.9Ω 0.001Ω 0.1Ω 1% of FS 1% of FS 2 100.0Ω to 250.0Ω 25.00Ω to 99.99Ω 0.1Ω 0.01Ω 1% of FS 1% of FS 1 10.00Ω to 25.00Ω 2.000Ω to 9.990Ω 0.01Ω 0.
Set Point Alarm Yes Input DO 6 Pin Round Display Display type Screen resolution Viewing area Backlit Dot matrix LCD with backlighting 110 x 128 68 x 74 mm Yes Other Data logging Data communication Auto Data logging GLP (Good Lab Practice) Ingress protection Weight 500 data sets IrDA / RS232C-Infrared Yes Yes IP 67 95mm (W) x 185mm (L) x 58.
7.
2-cell Electrode with ATC, cell constant K=10.0 35408-54 2-cell Electrode with ATC, cell constant K=0.1 35408-50 Galvanic Dissolved Oxygen probe with ATC with 10-ft cable 35640-50 Galvanic Dissolved Oxygen probe with ATC 25-ft cable 35640-52 Galvanic Dissolved Oxygen probe with ATC 50-ft cable 35640-54 Galvanic Dissolved Oxygen probe with ATC 100-ft cable 35640-56 Replacement batteries, AA. Pack of 4 09376-01 Replacement DO membranes, Pack of 5. 35640-74 Replacement DO membranes, Pack of 25.
damage during shipment, and insured against possible damage or loss. Eutech Instruments/ Oakton Instruments will not be responsible for any damage resulting from careless or insufficient packing. Warning: Shipping damage as a result of inadequate packaging is the user's/distributor’s responsibility. Please follow the guidelines below before shipment. 8.3 Guidelines for Returning Unit for Repair Use the original packaging material if possible when shipping the unit for repair.
Parameter Guide Temperature 1. Temperature Setup Mode 2. Temperature Calibration – ATC and MTC Mode 3.
36
1. Temperature Setup Temperature setup sub-group allows you to configure temperature measurement & calibration related settings of the meter. Temperature Setting Page Figure 13 : Temperature Settings Page Parameter Description Unit Sets the unit of measurement for temperature. Available units: ºC and ºF Sets the temperature compensation mode.
replacement probe. This procedure offers offset adjustment of the probe to ensure more accurate temperature measurement. Use a thermometer you are certain is accurate to measure the temperature of your sample. 2.2 Temperature Calibration for ATC mode Make sure you have selected ‘ATC’ and the required unit of measurement (ºC or ºF) in Temperature settings. Switch the meter on. Make sure the meter is in measurement mode. 1. Press CAL (F2) to go to calibration mode.
2.3 Temperature Calibration for MTC mode Make sure you have selected ‘MTC’ and required unit of measurement (ºC or ºF) in Temperature settings. Switch on the meter. Make sure the meter is in measurement mode. 1. Press CAL (F2) to go to calibration mode. 2. The meter shows the ‘Calibration-Rinse Electrode’ screen for a few seconds and then shows the cell constant adjustment screen. 3. Press NEXT (F3) key. (For conductivity mode only. In pH mode, pressing this key will take you to the Report page) 4.
40
Parameter Guide pH 1. pH Setup Mode 2. pH Calibration Mode 3.
42
1. pH Setup pH setup sub-group allows you to configure pH measurement & calibration related settings of the meter. The settings are displayed in 2 pages. Press NEXT-P (F2) and PREV-P (F1) to navigate through these pages. pH Settings Page 1 – pH Measurement & Calibration Figure 15: pH Settings Page 1 – pH Measurement & Calibration This page allows you to set pH measurement & calibration related settings of the meter.
pH Settings Page 2 – Calibration Due Alarm & pH Resolution Figure 16: pH Settings Page 2 – Cal-due Alarm & pH Resolution Parameter Description pH Calibration Due Specify number of days for the pH calibration alarm. The meter shows CAL DUE indicator after calibration due days are passed from the last calibration date. Available range: 0 to 30 pH Resolution Sets the resolution for pH measurement Available range: 0.1, 0.01, 0.001 Factory Default 5 Days 0.
Buffer Group USA NIST DIN PWB First Point of Calibration (when complete re-calibration) pH 7.00 pH 6.86 pH 6.79 pH 6.97 Other Available Calibration Points pH 1.68, 4.01, 10.01, and 12.45 pH 4.01, 6.86, 9.18, and 12.45 pH 1.09, 3.06, 4.65, 9.23, and 12.75 pH 4.10 Table 1: Buffer Groups and Calibration Points The meter automatically recognizes and calibrates to these standard pH buffer values, which makes pH calibration faster and easier. 2.
(a) (b) Figure 17: pH Calibration Screens 4. Rinse the electrode in clean water and prepare electrode for calibration. Note: If you wish to completely re-calibrate the meter, you need to clear previous calibration data. Press CLR-C (F2) key to clear previous calibration. The meter shows your confirmation screen. Press ENTER key to confirm deleting previous calibration. Once cleared, you need to calibrate the first point that corresponds to your selected buffer group. 5.
2.4 pH Calibration with a User-defined Buffer If you selected ‘USER’ (Custom) buffer in the pH Setup, the following screen is shown when you enter calibration mode. You need to prepare at least 2 custom buffers of known pH values. Note: Custom buffer solution values should be at least 1 pH unit apart from each other. Otherwise, the meter will not accept the buffer values. Figure 18 : User-define Buffer Screen 1. Rinse the electrode in clean water and prepare electrode for calibration.
ESC (F4) ENTER Exits from calibration and goes back to pH measurement mode Confirms the calibration Increases/decreases pH buffer reading (Not functional) 2.5 Calibration Report Calibration report gives you detailed information on calibration. It includes date & time, buffer group, offset, temperature, number of days calibration is over due and slope information. 2.5.1 To View Calibration Report: 1.
NEXT (F3) Goes to pH measurement mode ESC (F4) Goes to pH measurement mode ENTER (Not functional) (Not functional) 2.6 Average Slope Indicator of pH Probe Each time you perform pH calibration; the meter calculates the average slope of your probe and graphically indicates it in the header section of the LCD screen.
3.2 Indicators in pH measurement screen In pH measurement mode, the meter displays pH and temperature reading.
Parameter Guide ORP (mV) 1. mV Setup Mode 2. mV Calibration Mode 3.
52
1. mV Setup Mode There is no setup mode for mV parameter. 2. mV Calibration Mode 2.1 mV Calibration with a Standard ORP Calibration Solution Use a standard ORP solution of known value for calibration. You need to calibrate only 1-point. 1. Switch the meter on. Make sure the meter is in mV measurement mode. 2. Press CAL (F2) to go to calibration mode. Note: If the meter is password protected, you will be prompted to enter a password. 3.
7. Press ESC (F4) to go to measurement mode. The meter shows the relative mV reading of the solution in measurement mode, if offset is not zero. Note: The function of the keys available on the mV calibration/ report screen is same as previously described in pH calibration with a userdefined buffer section. 2.2 Calibration Report Calibration report gives you information on calibration. It includes date & time on which the last calibration was done and the offset. 2.2.1 To View Calibration Report: 1.
Item Description 1 Measurement mode indicator 2 Appears when the reading is stable 3 Appears when the reading is on hold 4 mV reading 5 Units of measurement 6 Temperature reading & units 7 Temperature compensation mode Note: If offset is not zero, the meter displays relative mV reading of the solution in measurement mode.
56
Parameter Guide Ion 1. Ion Setup Mode 2. Ion Calibration Mode 3.
58
1. Ion Setup Ion setup sub-group allows you to configure Ion measurement & calibration related settings of the meter. Ion Settings Page Figure 23: Ion Settings Page Parameter Description Ion unit Sets the unit of measurement for Ion. Available units: ppm, molar and mg/L Press (Up) or Factory Default ppm (Down) arrow key to change the value of a selected parameter. 2. Ion Calibration Mode 2.1 About Ion Calibration The meter supports Ion calibration up to 8 points with minimum of 2-points.
Use standard solutions of 0.001, 0.01, 0.1, 1, 10, 100, 1000 & 10000 ppm for calibration. You need to calibrate minimum of 2-points. See “For Ion” under “Preparing the Meter for Calibration” on page 18. 1. Switch on the meter. Make sure the meter is in Ion measurement mode. 2. Press CAL (F2) to go to calibration mode. Note: If the meter is password protected, you will be prompted to enter a password. 3.
7. The upper display shows the next calibration point. 8. Rinse the ISE with clean water. Place it in the next calibration solution and follow the steps 4 through 6 to calibrate other points. Note: The meter accepts the second and subsequent points only if the mV readings are within the calibration window (15mV/decade to 90mV/decade). 9. Press ESC (F4) to go to measurement mode.
3. Ion Measurement Mode In Ion measurement mode, the meter displays Ion concentration (in ppm, molar or mg/L) and mV reading. Depending on application, you can connect a suitable Ion Selective Electrode (ISE) to the meter. 3.1 Changing unit of measurement By default the measuring unit is ppm. You can select either molar or mg/L in the setup mode. 3.
Parameter Guide Conductivity 1. COND Setup Mode 2. COND Calibration Mode 3.
64
1. Conductivity Setup Conductivity setup screen present many options to control the operating parameters, which can be controlled and set from the conductivity setup screen. The settings are displayed in 2 pages. Press NEXT-P (F2) and PREV-P (F1) to navigate through these pages.
Note: 1. If pure water coefficient is set to ‘ENABLE’, the meter will apply pure water compensation automatically below 2 μS and switch back to linear compensation for above 2 µS. 2. If pure water coefficient is set to ‘DISABLE’, Lnr Temp Coefficient applies throughout the measurement. Conductivity Settings Page 2 – Hi-Lo Set Points Figure 28 : Conductivity Settings Page 2 – Hi-Lo Alarm This page allows you to set alarm limits for the conductivity measurement mode.
2. Conductivity Calibration Mode 2.1 Conductivity Calibration mode This option allows you to select automatic or manual calibration standard. In the automatic calibration mode, the meter automatically detects and verifies the appropriate known calibration standards solutions before accepting these particular calibration standards as one of its calibration values in a specific measurement range. Standard buffer values: 84.0 uS/ 1.413 mS/ 12.88 mS/ 111.
2.5 Pure Water Coefficient Pure water coefficient will be calculated and applied automatically for ultra pure water measurement if you select ‘ENABLE’ at appropriate set up pages. 2.6 To begin Calibration Make sure you have selected a standard solution with which you intend to perform conductivity calibration. Follow this procedure: 1. Switch on the meter and make sure the meter is in conductivity measurement mode. 2. Dip the probe into the calibration standard.
2.7 Manual Calibration Note: If you wish to completely re-calibrate the meter, you need to clear previous calibration data. Press CLR-C (F2) key to clear previous calibration. The meter shows you confirmation screen. Press ENTER key to confirm deleting previous calibration. Figure 31 : Conductivity Calibration Screen 1.
Figure 32 : Conductivity Calibration Screen – Auto 1. Rinse the electrode in de-ionized water and prepare electrode for calibration. Note: If you wish to completely re-calibrate the meter, you need to clear previous calibration data. Press CLR-C (F2) key to clear previous calibration. The meter shows you confirmation screen. Press ENTER key to confirm deleting previous calibration. 2. Dip the probe into the calibration standard. Immerse the probe tip beyond the upper steel band.
2.9 Calibration Report Calibration report gives you detailed information on calibration. It includes date & time, calibration points, range, factor and cell constant. 2.9.1 To View Calibration Report: 1. From conductivity measurement mode, press left or right arrow key to navigate to other available functions until you see REPO function in the LCD. 2. Press REPO (F2) key. The calibration report is shown in the display. 3. Press PRIN (F2) to transfer the calibration report to computer through IrDA.
3.
Parameter Guide Total Dissolved Solids (TDS) 1. TDS Setup Mode 2. TDS Calibration Mode 3.
74
1. TDS Setup The TDS setup screen presents many options to control the operating parameters, which can be controlled and set from the TDS setup screen. The settings are displayed in 2 pages. Press NEXT-P (F2) and PREV-P (F1) to navigate through these pages. TDS Settings Page 1 – Operating Parameters Figure 35 : TDS Settings Page 1 – Operating Parameters This page allows you to set operating parameters for TDS measurement mode: Factory Default ENABLE Parameter Description Pure water Coeff.
TDS Settings Page 2 – Hi-Lo Set Points Figure 36 : TDS Settings Page 2 – Hi-Lo Set Points This page allows you to set alarm limits for the TDS measurement mode: Parameter Alarm Set Points Hi ppm Lo ppt Description Set alarm limits for the TDS measurement mode: ENABLE – The measurement screen shows HI or LO alarm indicators when the meter reading is above ‘HI alarm value’ or below ‘LO alarm value’ specified in HI & LO ppm parameters (below) DISABLE – HI and LO alarms are not shown in the LCD Specify Hi
2. TDS Calibration Mode You can commence calibration in the TDS mode by using the conductivity calibration method and after setting the correct TDS factor. For the rest of the calibration process, follow steps as mentioned in the section on manual calibration. Use a standard solution of known value for calibration. Use the single-point standardization option to measure samples that are close in value and multi-point standardization to measure wide range of samples. 1. Switch the meter on.
Note: The function of the keys available on the TDS cell constant/calibration/ report screen is the same as previously described in the conductivity calibration section. 8. Rinse the probe with de-ionized water. Place it in the next solution and follow the steps 2 to 7 to calibrate other points, if ‘MULTI’ point calibration is selected. 9. Press ESC (F4) to go to measurement mode. The meter shows the TDS reading of the solution in measurement mode. 2.
3. TDS Measurement Mode In TDS measurement mode, the meter displays TDS and temperature reading. 3.
4. About TDS 4.1 Calculating TDS Conversion Factor You can calibrate your meter using TDS calibration standard solutions. The calibration standard only needs to give the TDS value at a standard temperature such as 25 °C. To determine the conductivity-to-TDS conversion factor use the following formula: Factor = Actual TDS ÷ Actual Conductivity @ 25 °C Definitions: Actual TDS: Value from the solution bottle label or as a standard you make using high purity water and precisely weighed salts.
Parameter Guide Salinity 1. Salinity Setup Mode 2. Salinity Calibration Mode 3.
82
1. Salinity Setup Salinity setup screen presents options to control the operating parameters, which can be controlled and set from the salinity setup screen. The settings are displayed in 2 pages. Press NEXT-P (F2) and PREV-P (F1) to navigate through these pages. Salinity Settings Page 1 – Operating Parameters Figure 39: Salinity Settings - Page 1 Factory Default Parameter Description Pure water Coeff.
Salinity Settings Page 2 – Hi-Lo Set Points Figure 40: Salinity Settings Page 2 – Hi-Lo Set Points This page allows you to set alarm limits for the salinity measurement mode: Parameter Alarm Set Points Hi ppm Lo ppt Description Set alarm limits for the salinity measurement mode: ENABLE – The measurement screen shows HI or LO alarm indicators when the meter reading is above ‘HI alarm value’ or below ‘LO alarm value’ specified in HI & LO ppm parameters (below) DISABLE – HI and LO alarms are not shown in t
3. The meter shows the ‘Calibration-Rinse Electrode’ screen for a few seconds and then shows the cell constant adjustment screen. Figure 41 : Salinity Cell Constant & Salinity Calibration 4. The upper display of the cell constant adjustment screen shows the cell constant of your cell and lower display shows the measured value of the solution with adjusted cell constant. Press & keys to adjust the cell constant for your custom solution in upper display. 5.
2.1 Salinity Calibration Report Calibration report gives you information on calibration. It includes date & time of the last calibration and the cal factor. 2.1.1 To View salinity Report: 1. From salinity measurement mode, press left or right arrow key to navigate to other available functions until you see REPO function in the LCD 2. Press REPO (F2) key. The calibration report is shown in the display. 3. Press PRIN (F2) to transfer the calibration report to the computer.
3. Salinity Measurement Mode In Salinity measurement mode, the meter displays salinity and temperature reading. 3.
88
Parameter Guide Resistivity 1. Resistivity Setup Mode 2. Resistivity Calibration Mode 3.
90
1. Resistivity Setup The resistivity setup screen presents many options to control the operating parameters, which can be controlled and set from the resistivity setup screen. The settings are displayed in 2 pages. Press NEXT-P (F2) and PREV-P (F1) to navigate through these pages.
Resistivity Settings Page 2 – Hi-Lo Set Points Figure 45: Resistivity Settings Page 2 – Hi-Lo Set Points This page allows you to select alarm limits for resistivity measurement mode: Parameter Alarm Set Points Hi Ω Lo Ω Press Description Set alarm limits for the resistivity measurement mode: ENABLE – The measurement screen shows HI or LO alarm indicators when the meter reading is above ‘HI alarm value’ or below ‘LO alarm value’ specified in HI & LO Ω parameters (below) DISABLE – HI and LO alarms are
2. Resistivity Calibration Mode Use a standard solution of known value for calibration. Use the single-point standardization option to measure samples that are close in value and multi-point standardization to measure wide range of samples. 1. Switch the meter on. Make sure the meter is in resistivity measurement mode. 2. Press CAL (F2) to go to calibration mode. 3. The meter shows the ‘Calibration-Rinse Electrode’ screen for a few seconds and then shows the cell constant adjustment screen.
Note: The function of the keys available on the resistivity cell constant/ calibration/ report screen is the same as previously described in conductivity calibration. 8. Rinse the probe with de-ionized water. Place it in the next solution and follow the steps 2 to 7 to calibrate other points, if ‘MULTI’ point calibration is selected. 9. Press ESC (F4) to go to measurement mode. The meter shows the resistivity reading of the solution in measurement mode. 2.
3. Resistivity Measurement Mode In Resistivity measurement mode, the meter displays resistivity and temperature reading. 3.
96
Parameter Guide Dissolved Oxygen Part A – % Saturation Mode 1. O2 % – DO Saturation Setup Mode 2. DO Calibration in % Saturation Mode (with ATC) 3. Percentage Saturation Measurement Mode Part B – Concentration Mode (mg/L)(ppm) 1. O2 % – DO Saturation Setup Mode 2. DO Calibration in (mg/L) or (ppm) Concentration Mode 3.
98
PART A - % Saturation Mode 1. O2 % - DO Saturation Setup The DO saturation (%) setup screen presents many options to control the operating parameters, which can be controlled and set from the DO setup screen. The settings are displayed in 2 pages. Press NEXT-P (F2) and PREVP (F1) to navigate through these pages.
DO (%) Settings Page 2 – Hi-Lo Set Points Figure 50: DO (%) Settings Page 2 – Hi-Lo Set Points This page allows you to set alarm limits for the DO saturation measurement mode. A visual alarm symbol will appear on the screen, if the DO value of the measurement is outside of the boundaries set by the minimum and maximum limits.
2. DO Calibration in % Saturation Mode (with ATC) 2.1 About DO(%) and DO (mg/L) Calibration The calibration of % Saturation of DO will linearly affect the measurement for DO in mg/L. The amount of oxygen dissolved in water will depend on its temperature, atmospheric pressure and its salinity. It is therefore very important that the temperature is calibrated if necessary prior to the DO calibration. Hence calibration in % Saturation of DO should be carried out first.
Figure 51: Saturation Calibration Screens 4. Rinse the probe well with de-ionized water. For best results, blot the end of the probe to dry it. Do not touch the membrane. 5. The meter then shows saturation calibration screen. The primary display will show the current value of measurement. Wait for the reading to stabilize. 6. Press ENTER key to confirm the calibration. The meter automatically calibrates to 100% air saturation and returns to the measurement mode.
CLR-C (F2) Clears previous calibration (if any) after ENTER key is pressed in confirmation screen NEXT (F3) Shows calibration report ESC (F4) Exits from calibration and goes back to saturation measurement mode ENTER Confirms the calibration (Not functional) 2.3 To calibrate 0% saturation Figure 53 : 0% Saturation Calibration Screen 1. Press MODE key to select % Saturation mode. 2. Immerse the DO probe in 0% solution. Stir gently to create a homogenous solution. 3.
2.3.1 % DO Calibration Report Calibration report gives you detailed information for the % saturation. It includes date & time, 0% and 100% saturation mV value, % saturation offset calibration and barometric pressure at which the solution was measured. 2.3.2 To View Calibration Report: 1. From % saturation measurement mode, press left or right arrow key to navigate to other available functions until you see REPO function in the LCD. 2. Press REPO (F2) key. The calibration report is shown in the display. 3.
2.5 Set barometer pressure range and barometric pressure units The PCD 650 meter is capable of measuring barometric pressure with its built-in pressure sensor. In the event that the pressure reading is inaccurate, you can calibrate the value from the setup menu. 2.6 Pressure compensation If pressure comp. is set to ‘ENABLE’, the meter will compensate for the barometric pressure at the location depending on the altitude.
3.
PART B – Concentration (mg/L) (ppm) Mode 1. O2 mg/L (ppm) – DO Concentration Setup DO Concentration (mg/L) setup screen present many options to control the operating parameters, which can be controlled and set from the DO setup screen. The settings are displayed in 2 pages. Press NEXT-P (F2) and PREVP (F1) to navigate through these pages.
DO (mg/L) – Page 2 Figure 57: DO (mg/L) – Page 2 This page allows you to set alarm limits for the DO concentration mode. A visual alarm symbol will appear on the screen, if the DO value of the measurement is outside of the boundaries set by the minimum and maximum limits.
2. DO Calibration in mg/L mode or ppm Concentration mode The amount of oxygen dissolved in a liquid will depend on its temperature, atmospheric pressure and salinity. It is therefore very important to set temperature and salinity correctly before attempting to do a calibration. The meter is capable of measuring barometric pressure with its built-in pressure sensor. In the event, the pressure reading is inaccurate, you can calibrate the value from the Setup menu. 1. Switch the meter on.
7. Press ENTER key to confirm the entered value. 8. Press ESC (F4) to go to measurement mode. The meter shows the concentration reading of the solution in measurement mode. Figure 59 : Concentration Calibration Note: The function of the keys available on the concentration calibration/ report screen is same as previously described in saturation calibration. 2.1 Concentration Calibration Report Calibration report gives you information on calibrated temperature and calibration at pressure and salinity.
pressure/oxygen concentration relationship of your sample. The salinity can be set from 0 to 50 ppt. 2.2.1 Auto Salinity Compensation If auto salinity comp. is enabled, the meter will automatically compensate for salinity for determining dissolved oxygen concentration of the sample. Auto salinity compensation will be applicable only in multi measurement mode and only if one of the measurement selected is conductivity related. This option is activated only in CD650 & PCD650.
5 Units of measurement 6 Temperature reading & units 7 Temperature compensation mode 8 HI & LO Alarm limits 9 Alarm indicator 10 ppt indicator for input of salinity value 11 DO calibration point 3.2 Dissolved Oxygen Probe 3.2.1 Dissolved Oxygen Principle The probe is a galvanic measuring element which produces an output proportional to the oxygen present in the medium in which it is placed.
Since any deposits on the membrane surface act as a barrier to oxygen diffusing through the membrane, the membrane must be cleaned at regular intervals to assure maximum reliability. After using the probe, rinse the probe in clean water and wipe it with a soft cloth or paper to avoid any hardening of deposits. If growth develops on the probe, use a disinfecting chemical to clean. NOTE: Although the membrane is strong and not easily damaged, wipe it gently while cleaning it.
To replace the membrane cap (with pre-installed membrane) 1. Replacement is much easier with single membrane housing. Unscrew the old membrane cap housing. 2. Hold the probe under hot running water and brush away the white oxide on the cylindrical anode with a stiff plastic brush – do not use metal cleaning material. 3. If the cathode has any deposits, remove them with a light scouring powder. Do not polish the cathode. 4.
3.2.4 Membrane/O-ring Replacement (Optional Procedure) It is recommended only experienced service personnel can perform this procedure. This procedure is OPTIONAL, and should only be performed if you have new membrane and O-ring. You are also required to have a membrane installation tool. These items are available as optional accessories in the “Accessories” section. 1. Pull off the probe guard. 2. Unscrew the membrane cap from the probe. 3.
Note: Membranes can only be used once. When a membrane cap is screwed onto the probe, the membrane is stretched by the cathode. If the same O-ring and the membrane is used a second time it will not fit perfectly onto the cathode. This will result in erratic readings. 3.2.5 Electrolyte Solution The electrolyte solution in your probe’s cap will deplete on usage and will need to be replaced periodically. Your new DO probe comes with accessories of one replacement electrolyte solution and a spare membrane cap.
NOTES 117
118
Note:The the TDS factory conversio default n factor is setting 0.50. for For more information on Eutech Instruments’/ Oakton Instruments’ products, contact your nearest distributor or visit our website listed below: Oakton Instruments 625 E Bunker Court, Vernon Hills, IL 60061, USA Tel: (1) 888-462-5866 Fax: (1) 847-247-2984 E-mail: info@4oakton.com Web-site: www.4oakton.com Eutech Instruments Pte Ltd.