Resistivity Settings Page 2 Hi-Lo Set Points
Figure 45: Resistivity Settings Page 2 Hi-Lo Set Points
This page allows you to select alarm limits for resistivity measurement mode:
Parameter Description Factory Default
Alarm Set
Points
Set alarm limits for the resistivity measurement mode:
ENABLE The measurement screen shows HI or LO
alarm indicators when the meter reading is above ‘HI
alarm value’ or below ‘LO alarm value’ specified in HI
& LO
parameters (below)
DISABLE HI and LO alarms are not shown in the
LCD
DISABLE
Hi
Specify Hi alarm value here. The Hi alarm occurs
when the resistivity reading goes above this value
(This parameter is not editable when ‘ALARM SET
POINT’ is disabled)
2.00
Lo
Specify Lo alarm value here. The Lo alarm occurs
when the resistivity reading goes below this value
(This parameter is not editable when ‘ALARM SET
POINT’ is disabled)
1.00
Press (Up) or (Down) arrow key to set high or low alarm point.