User Manual

107
PART B Concentration (mg/L) (ppm)
Mode
1. O2 mg/L (ppm) DO Concentration Setup
DO Concentration
(mg/L) setup screen present many options to control the
operating parameters, which can be controlled and set from the DO setup
screen. The settings are displayed in 2 pages. Press NEXT-P (F2) and PREV-
P (F1) to navigate through these pages.
DO (mg/L) Page 1
Figure 56 : DO (mg/L) Page 1
This page allows you to set operating parameters for the DO Concentration
measurement mode:
Parameter Function Factory
Default
Select DO Unit
Sets the unit of measurement for DO
Available units: mg/L and ppm
mg/L
Auto Salinity Comp
DISABLE No automatic compensate
ENABLEThe meter will apply measured salinity for DO
concentration of your sample.
Note: This option is activated only in CD650 & PCD650.
DISABLE
Set Salinity
Allows you to set the salinity (in ppt) for your solution
Available range: 0 to 50 ppt
0.0 ppt
NOTE: Auto salinity compensation will be applicable only in multi measurement
mode and only if one of the measurement selected is conductivity related. Otherwise,
a manually set salinity value will be applied.