User Manual

30
Resistivity
Range Measuring ranges Resolution Sub range Accuracy
5
4
3
2
1
10.00MΩ to 20.00MΩ
1.000MΩ to 9.999MΩ
500.0KΩ to 999.9KΩ
100.0KΩ to 500.0KΩ
10.00KΩ to 99.99KΩ
3.333KΩ to 9.999KΩ
1.000KΩ to 3.333KΩ
250.0Ω to 999.9Ω
100.0Ω to 250.0Ω
25.00Ω to 99.99Ω
10.00Ω to 25.00Ω
2.000Ω to 9.990Ω
0.01M
0.001M
0.1K
0.1
0.01
0.001
0.001
0.1
0.1
0.01
0.01
0.01Ω*
* Display resolution is 0.001 but actual
resolution shows 0.01.
1% of FS
1% of FS
1% of FS
1% of FS
1% of FS
1% of FS
1% of FS
1% of FS
1% of FS
1% of FS
1% of FS
1% of FS
Conductivity / TDS / Salinity /
Resistivity
COND 600 COND 610
Conductivity Range 0 to 200mS 0 to 500 mS
Salinity
-
80 ppt
Resistivity
-
0 to 20.00M
TDS 200 ppt (depending on TDS factor) 500 ppt (depending on TDS factor)
Conductivity Cell constant
0.010 to 10.000
Conductivity Cell 2 & 4 Cell
TDS Conversion Factor 0.400 to 1.000
Temperature Comp. Linear Linear & Pure
Cal-Auto/Manual Yes
Cal-Single/Multi Yes
Cal Due Alarm Yes (max-30 days)
Set Point Alarm Yes
Input Conductivity 8 Pin Round
Dissolved Oxygen DO 600, PD 650, CD 650, PCD 650
DO Range 0 to 600 %/ 90 mg/l
Resolution 0.1%, 0.01 mg/L
Accuracy ± 2%, ± 0.2 mg/L
Dissolved Oxygen Probe Galvanic
Mode % Sat, mg/L, ppm
Temperature Comp. Linear
Barometric Pressure Compensation Automatic
Barometric Pressure Range 450 to 825 mmHg
Barometric Pressure Resolution 1 mmHg
Barometric Pressure Accuracy ± 1%
Salinity Correction 0 to 50 ppt
Cal Due Alarm Yes (max -30 days)