Instruction Manual CyberScan pH 1500 Bench pH/mV Meter Technology Made Easy ...
PREFACE Thank you for selecting the Eutech Instruments CyberScan pH 1500 bench meter. The instruction manual serves to explain the use of the CyberScan pH 1500 bench meter as a step-by-step operational guide to help you familiarize with the meter’s features and functions. It is structured sequentially with illustration of diagrams that explains the various functions and setup menus available.
TABLE OF CONTENTS 1 INTRODUCTION 1.1 Introducing the CyberScan Series 2 GETTING STARTED 2.1 2.2 Meter Overview CONNECTORS 3 USING THE METER 3.1 3.2 3.3 3.4 Electrodes Temperature Probe Display / Keys Screen Display 4 SETUP MENUS 4.1 Using the Meter 4.1.1 P1.0 View pH Electrode Offset 4.1.2 P2.0 View pH Electrode Slope 4.1.3 P3.0 pH Buffer Group 4.1.4 P4.0 pH Resolution 4.1.5 P5.0 Temperature Unit 4.1.6 P6.0 Stability Indicator 4.1.7 P7.0 Baud Rate 4.1.8 P8.0 Parity Bit 4.1.9 P9.0 Stop Bit 4.1.10 P10.
Instruction Manual CyberScan pH1500 1 INTRODUCTION 1.1 Introducing the CyberScan Series Thank you for selecting the Eutech Instruments CyberScan pH1500 bench-top meter. This instruction manual describes the operation of the meter. The state-of-art meter that you have purchased is easy to operate and will guide you through the various functions by displaying easy to understand prompts.
Instruction Manual CyberScan pH1500 2 GETTING STARTED 2.
Instruction Manual CyberScan pH1500 2.2 CONNECTORS 1. Review the layout and arrangement of the rear connector panel.
Instruction Manual CyberScan pH1500 2. 3. Connect the electrode arm to the base. Connect the power cable to the connector cable to the rear connector panel power jack and to a power source.
Instruction Manual CyberScan pH1500 3 USING THE METER 3.1 Electrodes This meter allows you to use two types of electrodes: the conventional glass pH electrode and the Eutech Instruments field effect transistor (FET) pH electrode. If both types of pH electrodes are connected, the meter will read the FET electrode. 1. Carefully remove the protective cover from the end of the electrode.
Instruction Manual CyberScan pH1500 ) Do not discard the BNC shorting cap. 3. Rinse and blot-dry (don’t wipe) electrodes between each measurement. Rinse electrodes with distilled or deionised water, or a portion of the next solution to be measured. 4. ) Between measurements, store conventional pH electrodes storage solution, pH 4 buffer, or KCI solution. Always leave the filling hole of liquid filled combination electrodes open.
Instruction Manual CyberScan pH1500 3.2 Temperature Probe For more accurate pH measurement, a temperature probe (supplied with the meter) is used for Automatic Temperature Compensation (ATC). Once the temperature probe is connected into its socket (located at the back of meter), the ATC icon is displayed on the screen to indicate that the meter is in ATC Measure mode. If the temperature probe is not connected the meter displays 25.0 oC or 77 o F as the default temperature value.
Instruction Manual CyberScan pH1500 3.3 Display / Keys Overview of the meter screen display and function key layout. Press std key to initiate standardization. Or press std key at the Standardize mode allows you to exit and return to Measure mode. Press mode to select pH, mV and Relative mV. Press setup key to access setup for configuration of meter setting. Press print key to print stored data from memory or current reading to a PC or printer (depending on meter setup).
Instruction Manual CyberScan pH1500 3.4 Screen Display Familiarise yourself with the layout of the digital screen display.
Instruction Manual CyberScan pH1500 4 SETUP MENUS 4.1 Using the Meter The setup button brings you to the setup menus of the various parameters. Press setup while in measurement OR standardization to access to the setup menus. The ▲ and ▼ keys allow you to change several operating parameters. While in the setup mode you may: std setup print enter save view Press the std key at any time to return to the Measure screen without making a change or selection in the setup mode.
Instruction Manual CyberScan pH1500 In the setup mode, there are 12 setup menus for you to change the operating parameters to suit your individual preference as well as to configure RS232 communication interface protocol for printing. While you are in each parameter Measure mode, pressing setup key allows you to access all 12 setup menus (P1.0 to P12.0) in the pH mode and other setup menus for mV and Relative mV modes. Refer to Overview of Setup Menus below. ) Setup Menus Accessible pH mV REL mV P1.
Instruction Manual CyberScan pH1500 4.1.1 P1.0 View pH Electrode Offset This setup menu allows you to view the pH electrode offset in mV after 1-point standardization is successfully done. pH View press to set value press enter to accept ELECTRODE To view pH electrode offset 1. Access View the pH electrode offset menu from pH measurement screen by pressing setup key 2. Press ▲ and ▼ keys to scroll until display shown as above. view save 3. Press enter to view the offset.
Instruction Manual CyberScan pH1500 4.1.2 P2.0 View pH Electrode Slope This setup menu allows you to view the pH electrode slope in % after 2-point standardization is successfully done. pH % SLOPE View press to set value press enter to accept ELECTRODE To View pH Electrode Slope 1. Access the View pH Electrode Slope menu from the pH Measure screen by pressing the setup key. 2. Press ▲ and ▼ keys to scroll until display shown as above. view save 3. Press enter to view the offset.
Instruction Manual CyberScan pH1500 4.1.3 P3.0 pH Buffer Group This setup option allows you to select from 3 different buffer groups, each containing 5 buffers, for auto buffer recognition. The 3 buffer groups are: USA buffers: European buffers: NIST buffers: 2, 4, 7, 10 and 12 1, 3, 6, 8 and 10 1.68, 4.01, 6.86, 9.18 and 12.46 To Set pH Buffer Group 1. Access the Select Buffer menu from the pH Setup Screen by pressing the setup key.
Instruction Manual CyberScan pH1500 3. Press enter to make a selection.. BUFFER select 2 4 7 10 view save BUFFER select 1.68 4.01 6.86 9.18 12.46 view save 1 3 12 6 8 10 13 BUFFER select print enter 4. To make a change or selection to the desired buffer group, press enter key. 5. Use the ▲ or ▼ keys to scroll through the options available until the desired buffer group is displayed. 6.
Instruction Manual CyberScan pH1500 4.1.4 P4.0 pH Resolution This setup option allows you to select various pH resolution either 0.1, 0.01 or 0.001. To Select pH Resolution 1. Access the Select pH Resolution menu from the pH Measure screen by pressing the setup key. 2. Use the ▲ or ▼ key to scroll until the display “rESoL” is displayed.
Instruction Manual CyberScan pH1500 3. Press enter key to make selection and use the ▲ or ▼ key to choose either 0.1 (one decimal place), 0.01 (two decimal place) or 0.001 (three decimal place). pH save view save view pH pH print enter std 4. Press enter key to confirm selection OR press std to exit from this page. 5. Continue to access other setup menu using ▲ or ▼ keys OR press std key to exit from the setup and return to Measure mode.
Instruction Manual CyberScan pH1500 4.1.5 P5.0 Temperature Unit This setup option allows you to select unit of measure for Temperature either in °C or °F. To Select Temperature Unit 1. Access the Select Temperature Unit menu from any parameter Measure screen by pressing the setup key. 2. Use the ▲ or ▼ key to scroll until the Select Temp Unit icon is displayed.
Instruction Manual CyberScan pH1500 3. Press enter key to make selection and use the ▲ or ▼ key to choose either °C or °F. sel ect °C view save sel ect °F print enter std 4. Press enter key to confirm selection OR press std to exit from this page. 5. Continue to access other setup menu using ▲ or ▼ keys OR press std key to exit from the setup and return to Measure mode.
Instruction Manual CyberScan pH1500 4.1.6 P6.0 Stability Indicator This setup option allows you to set the stability indicator to be displayed on the screen whenever reading has stabilized, thus minimizes guesswork. To Set Stability Indicator 1. Access the Set Stability Indicator menu from any Measure screen by pressing the setup key. 2. Use the ▲ or ▼ key to scroll until the screen displays the stable icon is shown. view save STABLE press to set value press enter to accept print enter 3.
Instruction Manual CyberScan pH1500 4.1.7 P7.0 Baud Rate This setup option allows you to set the baud rate (bits per second) of the communication protocol interface. To Set Baud Rate Print press to set value press enter to accept 1. Access the Set Baud Rate menu from any Measure screen by pressing the setup key. 2. Use the ▲ or ▼ key to scroll until the screen display as shown.
Instruction Manual CyberScan pH1500 3. Press enter key to make selection and use the ▲ or ▼ key to toggle between 4800, 9600, 19200 and 38400. view save Print Print press toset setvalue value press to enter to press enter press to accept accept 4. Press enter key to confirm selection OR press std to exit from this page. 5. Continue to access other setup menu using ▲ or ▼ keys OR press std key to exit from the setup and return to Measure mode.
Instruction Manual CyberScan pH1500 4.1.8 P8.0 Parity Bit This setup option allows you to set the parity bit of the communication protocol interface. To Set Parity Bit 1. Access the Set Parity Bit menu from any Measure screen by pressing the setup key. 2. Use the ▲ or ▼ key to scroll until the screen display as shown.
Instruction Manual CyberScan pH1500 3. Press enter key to make selection and use the ▲ or ▼ key to toggle between 0 (none), 1 (odd) or 2 (even). save view save view 4. Press enter key to confirm selection OR press std to exit from this page. 5. Continue to access other setup menu using ▲ or ▼ keys OR press std key to exit from the setup and return to Measure mode.
Instruction Manual CyberScan pH1500 4.1.9 P9.0 Stop Bit This setup option allows you to set the stop bit of the communication protocol interface. To Set Stop Bit 1. Access the Set Stop Bit menu from any Measure screen by pressing the setup key. 2. Use the ▲ or ▼ key to scroll until the screen display as shown.
Instruction Manual CyberScan pH1500 3. Press enter key to make selection and use the ▲ or ▼ key to toggle between 1 or 2. save view save view 4. Press enter key to confirm selection OR press std to exit from this page. 5. Continue to access other setup menu using ▲ or ▼ keys OR press std key to exit from the setup and return to Measure mode.
Instruction Manual CyberScan pH1500 4.1.10 P10.0 Print Data This setup option allows you to print current displayed data or data stored in the meter’s memory to a computer or printer via its RS232 interface port. Note all the communication protocol for both the meter and computer/printer must match before successful printing can be performed. To Select Print Data Option (Current / Memory) 1. Access the Select Print Data Option menu from any Measure screen by pressing the setup key. 2.
Instruction Manual CyberScan pH1500 4.1.11 P11.0 Clear Stored Data This setup option allows you to clear all stored data sets (from previous measurements) in the meter’s memory for new data to be stored. Note old data sets will be overwritten by any new data sets in the event when the stored locations have exceeded. To Clear Stored Data 1. Access the Clear Stored Data menu from any Measure screen by pressing the setup key. 2. Use the ▲ or ▼ key to scroll until the screen displays as shown.
Instruction Manual CyberScan pH1500 3. Press enter key to make selection and use the ▲ or ▼ key to toggle between NO or YES. Select YES to clear all stored data sets or NO if no change is made. view save print enter 4. Press enter key to confirm selection OR press std to exit from this page. std 5. Continue to access other setup menu using ▲ or ▼ keys OR press std key to exit from the setup and return to Measure mode.
Instruction Manual CyberScan pH1500 4.1.12 P12.0 Clear Buffer Values This setup option allows you to clear the standardized buffer values. To Clear Buffer Values 1. Access the Clear Buffer Values menu from the pH Measure screen by pressing the setup key. 2. Use the ▲ or ▼ key to scroll until the screen displays the clear BUFFER icon along with the previously buffer values being standardized. view save clear BUFFER press to set value press enter to accept print enter std 3.
Instruction Manual CyberScan pH1500 4.1.13 P13.0 View mV offset This setup option allows you to view the standardized mV offset values. To view mV offset values 1. Access the View mV offset values menu from the REL mV Measure screen by pressing the setup key. 2.
Instruction Manual CyberScan pH1500 4.1.14 P14.0 Clear mV offset This setup option allows you to clear the standardized mV offset values. To Clear mV offset Values 1. Access the Clear mV offset values menu from the REL mV Measure screen by pressing the setup key. 2. Use the ▲ or ▼ key to scroll until the screen displays the Offset screen with the CLEAR icon along. 3. Press enter key. 4. Use the ▲ or ▼ key to scroll between YES or NO to clear the REL mV offset values. 5.
Instruction Manual CyberScan pH1500 5 PH OPERATION 5.1 Standardization Because electrodes vary in their response, you must standardize or calibrate your pH meter and electrode to compensate for electrode variation. The more frequently you standardize or calibrate, the more accurate your measurements. Standardize at least daily, or more often, for accurate results.
Instruction Manual CyberScan pH1500 Press std key to begin standardization. Similarly pressing std key again at the Standardize mode exits and returns to the Measure mode. Press mode key to toggle between different modes until the pH Measure mode is displayed. Immerse the rinsed electrode(s) into a buffer from the selected buffer group. Stir moderately. ) This meter allows for automatic calibration using up to 6 buffers.
Instruction Manual CyberScan pH1500 A. Using Standard Buffer Group a. Press std key to access the standardize mode while you are in the pH Measure mode. b. Meter starts scanning the buffer from the already selected buffer group. Proceed to step c to begin standardization. Press setup key if you need to select other buffer group. Use ▲ or ▼ key to select desired buffer group and press enter key to confirm selection. BUFFER select 2 4 7 10 view save BUFFER select 1.68 4.01 6.86 9.18 12.
Instruction Manual CyberScan pH1500 f. Repeat the above step c through e for the second and subsequent calibration points. Meter automatically returns to the Measure mode after the last 5th calibration point has been performed. In the pH Measure mode, the screen shows all the buffer values being standardized. An error message Err2 will be displayed if there is more than 6 points pH standardization being performed.
Instruction Manual CyberScan pH1500 B. Using Custom Buffer Values std a. Press std to access the standardize mode while you are in the pH Measure mode. b. Proceed to step c to begin standardization. Press setup key if you need to select other buffer group. Use ▲ or ▼ key to select desired buffer group and press enter key to confirm selection. c. The meter automatically scans and recognizes the pH buffer being standardized; it locks and displays the closest suitable pH buffer value on the screen.
Instruction Manual CyberScan pH1500 5.2 pH Electrode Diagnosis Indicator There is an electrode diagnosis icon that indicates the condition of pH electrode being used. The STABLE icon will appear once the reading is stabilised. If the meter is calibrated with 2 or more points, then meter displays the electrode characteristic indicator bar with appropriate segments.
Instruction Manual CyberScan pH1500 6 mV OR RELATIVE mV OPERATION 6.1 Absolute mV Millivolts (mV) measurement is used primarily for measuring redox potential (also known as ORP, oxidation reduction potential). You will normally use a platinum ORP electrode, combined with a reference electrode, to measure redox potential (ORP) of a test sample. ORP measurements indicate the oxidizing or reducing capability of a solution.
Instruction Manual CyberScan pH1500 6.2 Relative mV The Relative mV mode helps you to standardize certain analytical and monitoring activities such as filtrations. mode std 1. Press mode until your meter displays the Relative mV mode. 2. Immerse the electrode in a sample solution. 3. Press std and use the or keys to adjust the mV value to the std solution value. Press enter key to confirm the adjustment. Note: Maximum adjustable mV offset value is ±150.0 mV.
Instruction Manual CyberScan pH1500 6.3 Temperature Calibration 1. Press mode to go to the absolute mV measure screen 2. Press std. this will take you to the temperature calibration (MTC or ATC) 3. Press p or q to adjust the temperature. 4. Press enter to confirm the reading. Note: Maximum adjustable temperature offset is ±5°C/ 9°F.
Instruction Manual CyberScan pH1500 7 MEASUREMENT 1. Immerse the electrode (along with temperature probe if available for the measurement of pH) into the sample solution. Stir moderately. Make sure that the meter is in the Measure mode. 2. When the meter senses that the reading has stabilized, the stable icon will appear under the reading. The reading may be recorded at this time. ) Stirring with a magnetic stir bar and stirrer provides faster electrode response.
Instruction Manual CyberScan pH1500 8 MEMORY 8.1 Store Value Into Memory save In any Measure mode, press ▲ save key to store the displayed reading into the meter’s non-volatile memory. A memory location is shown momentarily and the meter returns to the Measure mode. 8.2 Recall Value From Memory print enter save view In any Measure mode, pressing ▼ view key retrieves data from the meter’s memory on the Last-In-First-Out (LIFO) basis. The screen displays the last stored memory location.
Instruction Manual CyberScan pH1500 9 PRINT DATA 9.1 Printing Data print enter std Depending on the print option in the meter setup, pressing print key allows you to print either current displayed reading or stored data from meter to a PC or printing device via a RS232 communication cable. Ensure that both meter and peripheral have the same configuration in terms of baud rate, parity bit and stop bit.
Instruction Manual CyberScan pH1500 10 pH THEORY The measurement of pH plays an important role in quantifying and controlling acidity and alkalinity levels for industry and research. pH is a measure of the acidity or alkalinity of a solution and can be represented by this equation: pH= -log [H+] With [H+] representing the concentration of hydrogen ions in the solution. pH is sometimes referred to as power of the hydrogen ion in a solution.
Instruction Manual CyberScan pH1500 10.1 Measuring pH To measure pH, the meter receives a millivolts signal from a glass bulb electrode that is sensitive to hydrogen ions. Therefore the potential developed at the glass bulb is directly related to the pH of the solution. The glass bulb electrode is paired with a reference electrode which completes the electrical measuring circuit and provides a stable reference point.
Instruction Manual CyberScan pH1500 11 CLEANING This meter requires no regular maintenance, but it is recommended to occasionally wipe down the front with a damp cloth from time to time. If there are any further questions regarding maintenance, call the Eutech Instruments Technical Support hotline at (+65) 6778 6876. 12 TROUBLESHOOTING Most troubles arise out of a problem with the electrode, not the meter. However, a meter check can be performed, and is simple to do. 1.
Instruction Manual CyberScan pH1500 13 METER SPECIFICATIONS Display Screen Size Measurement display height Temp./others display height Keypad controls Custom LCD 7.2 x 10.6 cm 5.5 cm 1.6 cm 7 keys (membrane) Memory Internal Diagnostics yes pH Mode Range Resolution Relative Accuracy Automatic Buffer Recognition Calibration Points FET -1.999 to 19.999 0.1/0.01/0.001 ± 0.002 yes 5/6 yes mV Mode Range Resolution Accuracy -1800.0 to 1800.0 0.1 ± 0.
Instruction Manual CyberScan pH1500 14 ACCESSORIES Consult your Authorized Distributors for these items and other range of specialized pH electrodes or Ion Selective Electrodes.
Instruction Manual CyberScan pH1500 Calibration Solutions EC-BU-1BT pH 1.68 calibration buffer (480ml bottle) EC-BU-4BT pH 4.01 calibration buffer (480ml bottle) EC-BU-7BT pH 7.00 calibration buffer (480ml bottle) EC-BU-10BT pH 10.01 calibration buffer (480ml bottle) EC-BU-12BT pH 12.
Instruction Manual CyberScan pH1500 15 WARRANTY Eutech Instruments supplies this bench meter with a 3-year warranty and 6-month warranty for electrode against manufacturing defects from the date of purchase. If repair or adjustment is necessary and has not been the result of abuse or misuse within the warranty period, please return, freight prepaid, and correction will be made without charge. Out of warranty items will be repaired on a charge basis.
Instruction Manual CyberScan pH1500 16 NOTICE OF COMPLIANCE Warning This meter generates, uses, and can radiate radio frequency energy. If not installed and used properly, that is in strict accordance with the manufacturer’s instructions, it may cause interference to radio communications.
For more information on Eutech Instruments products, contact your nearest Eutech Instruments distributor or visit our website listed below: Manufactured by: Eutech Instruments Pte Ltd. Blk 55, Ayer Rajah Crescent, #04-16/24 Singapore 139949 Tel: (65) 6778 6876 Fax: (65) 6773 0863 E-mail: marketing@eutechinst.com Web-site: www.eutechinst.