Operating Manual
Chapter 8 Serial Ports, Contact Inputs & Special Fcts Accu-Wave 
8-10 TN Technologies 
can be formatted for various terminals (CRTs), printer presentations, or for host 
computers. 
Data streaming on a party-line (RS-485 multiple units) is automatically suspended when 
you send the “SLEEP” command, and is normally suspended when you “CONNECT” to 
any unit on the party line. You can tell a unit to “output while in connect” using the “Set 
up Port …” menus (described below) so the unit will continue to data stream, except while 
you are in the menu system. A unit with a unit number of zero will automatically “output 
while in connect.” 
A reading is a measurement such as level or volume. A data set is all of the readings being 
sent by a particular unit. By default, a NEWLINE is sent after each reading. A HOME 
and a CLEAR are sent after each set. This causes the data to update at the same place on 
the screen. 
Set up Port 1 RS232 (Port 2 RS485) Data Transmission 
This menu controls the selection, formatting, and transmission of real-time measurement 
data to a serial device via the RS-232 (RS-485) ports. 
“Modify Port 1 RS232 (Port 2 RS485) Configuration” Menu 
Display  Comments 
Set up port 1 RS232 
data transmission→→ 
(data streaming) 
 NEXT↓↓ 
Press → key to access the “Set up port…” menus to 
configure data transmission (data streaming). 
←←Exit from: RS232 
data transmit setup. 
Continue↓↓ 
 Data Transmit HELP→→ 
Press → to access the “Data Transmit Help” screens. 
Press ↓ to continue with setup. 
Do not inhibit RS232 
data streaming. 
Change to “Do”→→ 
Continue as is.↓↓ 
Turns data streaming on or off. If you inhibit data 
streaming (select “Do”), data transmission stops until 
you again select “Do not.” 