Party–Line Communications 
This section describes various aspects of setting up and using party-line communications 
for multiple gauges. 
Setting up a Party-Line 
To set up party-line communications from the local keypad and display of each gauge on 
the party-line, use the following procedure. 
1. Wire all units to the RS-485 common lines as per the wiring instructions in the 
Accu-Wave Installation and Maintenance Manual provided with your gauge. 
2. For each unit, ensure the “Modify port …” menu communication settings is the 
same (baud rate, parity, and handshake). Assign each gauge a unique non–zero 
unit number. See page 8-8. 
3.  In the “Set up port … data tranmission” menu (page 8-10), set up the desired RS-
485 measurement selections, data formats, user messages, etc., for each unit.  
After completing the data transmission setup, write down the “must be >=” value 
displayed in the “window code” menu item. This is the minimum window code 
required to transmit the data set for a particular gauge. 
4.  Using the “Set up port … data transmission” menu, configure the “master” unit 
as follows. (The master unit is usually assigned the lowest unit number, typically 
unit #1.) 
− Specify that this unit is the master unit. 
− Set the window code to be greater than or equal to the largest of the “must be 
>=” window code values found for all of the units in step 3. 
− Specify an appropriate RS-485 update time (see the next section, “Master 
Update Time.” 
5. Using the “Set up port … data transmission” menu, configure each of the “slave” 
units as follows: 
− Specify that the unit is a slave. 
− Specify the same (largest) value for the window code as used for the master 
unit in step 4. 
− Specify that the RS-485 update time is zero (0). 