Operating Manual
Accu-Wave Chapter 8 Serial Ports, Contact Inputs & Special Fcts 
TN Technologies 8-17 
When a unit is in “Connect” mode, the unit’s setup menus can be accessed (if not in Blind 
or Printer mode) and the unit will continue to send updates if “Do data stream in connect 
mode” has been enabled. Connect mode is exited when a valid DISCONNECT command 
is received. 
A unit in Sleep mode will not respond to anything except a valid WAKEUP Command. 
Party-Line Commands 
Some useful escape codes are: 
ESC [ Z# uu CONNECT Command. (uu = unit number, e.g., 01).  
ESC O Q DISCONNECT Command  
ESC X C SLEEP Command (ANSI) 
ESC X D WAKEUP Command 
ESC X 1 CONNECT Acknowledge (gauge sends this to acknowledge the 
CONNECT command) 
ESC X 2 Acknowledge DISCONNECT Command 
ESC O V ID REQUEST Command 
ESC X 4 ID REQUEST TERMINATED Command 
ESC [Yuu;cc;vvdd  POLLING Command,  
where 
 uu = unit number (in hex) on party line 
 cc = command (in hex) 
 vv = unit number + 3 (in hex) 
 dd = command number +3 (in hex) 
Presently, only three “cc” command values are supported: 
86 – Single Data Stream Update on RS-485 
8B – Download NVRAM contents to PC on RS-485 
8D – Upload NVRAM contents to gauge 
Example: 
ESC [Y01;86;0489 sends the “Single Update” command to Unit 1.  
Party-Line Limitations 
The RS-485 party-line uses half-duplex communications. That is, only one system can 
send information at any given time. For instance, if you are entering data during the time 
window for a gauge to send its output, garbled characters may result. This occurs mainly 
when the gauge’s menu system is being accessed in Connect mode. In this circumstance, 
you should verify that the gauge received the correct setup information before continuing. 