Operating Manual
Accu-Wave Chapter 9 Security, Service & Diagnostic 
TN Technologies 9-7 
“Snapshot” Menu (cont.) 
[Service only item] 
Display  Comments 
base 
signal strength 
113.9 
 NEXT↓↓ 
Signal Strength of “best” candidate echo as reported 
in the echo report. 
weighted 
signal strengh 
113.9 
 NEXT↓↓ 
Signal strength of “best” candidate echo after 
weighting factors have been applied. 
View echos being 
processed→→ 
 NEXT↓↓ 
Allows you to examine the echo report for each echo 
within the range of interest. Press → to access the 
echo reports, then press ↓ to scroll through the echo 
reports. The echo report is described on page 5-11. 
View marker report→→ 
 NEXT↓↓ 
Reports relative control levels at which RF frequency 
markers occur. See “View Marker Report” menu table 
below. 
DSP restart count 
0 
 NEXT↓↓ 
Internal value of 
Sensor signal 
1.871E4 
←←CONT UPDATE→→ NEXT↓↓ 
Filtered value of echo position after electrical offset is 
subtracted. Press ← for continuous measurement 
updates. Press ← again to return to FREEZE mode. 
Internal value of 
distance 
441.0 cm 
←←CONT UPDATE→→ 
Filtered value of echo position (distance) converted to 
centimeters. Most internal values in the software are 
stored in CGS units (centimeters, grams, seconds). 
Note: The next three menu items will only display meaningful values when a 
single current output board is installed. 
Internal value of 
I out % 
60.80 % 
←←CONT UPDATE→→ 
Internal value of current output - % of range. 
Internal value of 
I out flt 
5429 (8000=100%) 
←←CONT UPDATE→→ 
Internal value of 
I out fix 
5428 (8000=100%) 
UPDATE→→ 
This is the value actually sent to the current output 
board. 










