Operating Manual
Accu-Wave Chapter 9 Security, Service & Diagnostic 
TN Technologies 9-9 
User Service & Related Items 
“User Service & Related Items” Menu  
Display  Comments 
System restart. Does 
not affect user data 
←←Exit this menu 
 NEXT↓↓ EXECUTE CMD→→ 
Re-initializes system including the DSP. Erases 
temporary memory, but does not affect user-entered 
setup data. 
Erase all entries !!! 
(except COMM setup) 
←←Exit this menu 
 NEXT↓↓ EXECUTE CMD→→ 
Erase all previously entered set up data. All settings 
except for serial communication settings are reset to 
factory defaults. Gauge will behave just like the first 
time power was applied. 
Clear all holds 
(none now in effect) 
←←Exit this menu. 
  NEXT↓↓ EXECUTE CMD→→ 
Execute this command to clear all holds (if any holds 
are in effect). 
Hold current output 
at max (20.00 mA) 
←←Exit this menu. 
 NEXT↓↓ EXECUTE CMD→→ 
Holds all current outputs at the value entered for the 
maximum current output. 
Hold current output 
at min (4.000 mA) 
←←Exit this menu. 
 NEXT↓↓ EXECUTE CMD→→ 
Holds all current outputs at the value entered for the 
minimum current output. 
Hold current output 
at zero mA 
←←Exit this menu. 
 NEXT↓↓ EXECUTE CMD→→ 
Holds all current outputs at 0.0 mA. 
Current output 
hold mode value 
50.00% of scale 
 NEXT↓↓ 
Enter value in % for the mid-range current output 
hold value. Default value is 50.00%. 
Hold current output 
at 50.00% of scale 
←←Exit this menu. 
 NEXT↓↓ EXECUTE CMD→→ 
Press → to execute command to hold current output at 
the mid-range value specified in the previous menu 
item. 