Operating Manual
Accu-Wave Chapter 9 Security, Service & Diagnostic 
TN Technologies 9-15 
Signal Diagnostics (Factory Service Submenu) 
“Signal Diagnostics” Menu 
[Service Only Item] 
Display  Comments 
Do not spec to sweep 
sig for scope 
Change to “Do”→→ 
Continue as is.↓↓ 
Time shares the sweep control line to the head with 
the frequency spectrum for use with a scope. A low 
going pulse on the line is provided for scope trigger. 
Do not I.F. to sweep 
sig for scope 
Change to “Do”→→ 
Continue as is.↓↓ 
Time shares the sweep control line to the head with 
the I.F. (audio) signal date for use with a scope. A 
low going pulse on the line is provided for scope 
trigger. 
Set gain = 0.000 
(1.0 – 40, overrides 
automatic value) 
(1.252 in use) NEXT↓↓ 
Normally, the preamp gain is adjusted based on the 
dielectric constant and the measurement range. This 
selection disables the range determined portion of the 
gain. 
Dielectric constant 
of process 0.000 
(presently 80.00) 
 NEXT↓↓ 
If value is 0.00, the “presently” assumed value is 
used. Enter a non-zero value to change default. 
Dielectric constant determines gain applied to 
microwave signal. 
Do not disable 
range-GAIN function. 
Change to “Do”→→ 
Continue as is.↓↓ 
gain set step 1 
set=20.00 gain=1.252 
 NEXT↓↓ 
Gain is a non-linear function of gain setting. 
Select step number. 
+ gain set 1 steps 
set=20.00 gain=1.252 
←←Reject command 
 EXECUTE CMD→→ 
- gain set 1 steps 
set=20.00 gain=1.252 
←←Reject command 
 EXECUTE CMD→→ 
track MIN & MAX on 
measurement 0 
none 
 NEXT↓↓ CHANGE→→ 
Press → to select measurement number. Selecting a 
non-zero measurement number will cause additional 
menu items to be displayed allowing you to track the 
min and max values of the selected measurement. 