Operating Manual
Accu-Wave Chapter 3 Set up Level, Level Alarms, and Volume 
TN Technologies 3-1 
Chapter 3 Set up Level, Level Alarms, and Volume 
When power is first applied, the Accu-Wave gauge will be configured to measure distance 
with a maximum range of 112 ft (34.1 m). Until configured otherwise, the gauge 
continuously measures and displays distance and signal strength, except when you are in a 
menu.  
Note:  Allow 2 minutes after start up for the readings to stabilize. 
The measurement display, shown in Figure 3.1, shows the distance to the strongest echo in 
feet and the amplitude or signal strength (sig str) of the echo. 
23.69 ft 
 114 sig str 
For setup press→ 
Figure 3.1 Measurement Display 
Press → to begin setting up the primary measurement. The gauge continuously makes 
measurements, even while in the menu display. The primary measurement is distance by 
default, but you can use the “Set up Level,...” menu to change to level, as described later 
in this chapter. 
Note: The appearance of the menu items as described in this manual will often 
vary slightly from the actual menu item displayed on your gauge. The 
appearance of many menu items (words and numbers) varies dynamically 
with context and will depend on the parameter values and selections you 
enter during setup. 
Note: After you have configured the gauge and the configuration settings have 
been saved, the gauge retains the configuration settings even when power 
is turned off. 
Typically the displayed distance will correspond to the echo from the surface of the 
process material as desired, but in some cases the strongest echo will be from another 
cause, such as: 
− a multiple bounce from the process material and the top of the vessel. This 
multiple bounce might cause a warning message, such as “signal saturated.” 
− a multiple bounce from a nozzle obstruction. 
− an echo from an obstruction within the vessel. 
If it appears that the distance displayed on your gauge is incorrect, you can later use the 
radar fine tuning tools (described in Chapter 5) to reduce the amplitude of the unwanted 
echoes so that the system locks onto the correct echo. Before using these tools, you must 
complete the setup for the primary measurement and define the correct range of interest, as 
described in the next section. 










