Operating Manual
Accu-Wave Chapter 3 Set up Level, Level Alarms, and Volume 
TN Technologies 3-7 
Alarms: Set Point, Clear Point, and Dead Band 
An alarm is defined with either a set point - clear point configuration, as shown in Figure 
3.4, or a set point - dead band configuration, as shown in Figure 3.5. 
The set point defines the measurement value, either level or distance for the primary 
measurement, at which the alarm is activated. The clear point or the dead band defines the 
measurement value at which the alarm is cleared (alarm ceases). 
A clear point sets a fixed measurement value at which the alarm clears. The position of the 
clear point is independent of the set point and remains the same even if the set point is 
moved. As illustrated in Figure 3.4, changing the set point from 3.0 m to 3.5 m has no 
effect on the clear point which remains at 2.0 m. 
Figure 3.4 Set Point / Clear Point Alarm 
A dead band defines a fixed distance between the set point and an implicit clear point. If 
the set point is moved, the implicit clear point moves also, maintaining the distance from 
the set point specified by the dead band. As illustrated in Figure 3.5, changing the set point 
from 2 m to 3 m moves the implied clear point from 4 m to 5 m. The relative distance 
between the implied clear point and the set point remains fixed at 2 m, the dead band 
value. 
If you may want to later change the alarm set point without affecting the point where the 
alarm is cleared, use a clear point configuration. Alternately, if you want the alarm clear 
point to be defined at a fixed distance relative to the set point, use a dead band 
configuration. 
b. Modified Set 
Point /Clear Point 
a. Initial Set Point 
/Clear Point Alarm 
3.0 m
Clear Point 
New 
Set Point 
Clear Point 
Initial 
Set Point 
2.0 m
3.5 m










