Operating Manual
Accu-Wave Chapter 3 Set up Level, Level Alarms, and Volume 
TN Technologies 3-9 
“Set Up Alarm 1” Menu (Primary Measurement) 
This table describes the menu items in the “Set up Alarm” menu group. 
“Set up Alarm” Menu 
Display  Comments 
Set up alarm 1 
(Alarm point, etc.)→→ 
  NEXT↓↓ 
Menu is normally available only if relays are 
installed. Press → to access alarm menu or ↓ to skip 
alarm set up. After alarm 1 is set up, menu to set up 
alarm 2 will be available, and so on. 
Note: The appearance of the following menu items depends on whether distance 
or level has been selected as the primary measurement. Menu items are 
shown for distance. Differences are noted in the Comments. 
 You must enter a set point to activate the rest of the Set up Alarm menu. 
←←Exit alarm 1 setup 
Alarm 1 set point 
0.000 ft dist 
 NEXT↓↓ HELP→→ 
Enter the distance (or level) from the zero distance 
(or zero level) point to the set point; the point at 
which the alarm is activated. The zero point value 
was entered during the primary measurement set up. 
Alarm 1 clear based  
on clr point 
Chng to “dead band”→→ 
Continue as is.↓↓ 
Select a clear point or dead band configuration. A 
clear point is the exact measurement value at which 
the alarm is cleared. A dead band specifies the span 
between the set point and the implicit clear point. 
Alarm 1 clear point 
0.000 ft dist 
{Makes alarm “Low” 
limit} NEXT↓↓ HELP→→ 
Enter desired clear point value: the distance (or level) 
relative to the zero distance (or zero level) point. If 
dead band is selected, enter the dead band span 
relative to the set point. A negative dead band makes 
the alarm high (set pt > clear pt. A positive dead band 
makes the alarm low (set pt < clear pt). 
Alarm 1: ft dist  
is indicated by 
controlling relay 1 
 NEXT↓↓ CHANGE→→ 
Select the action used to indicate that alarm 1 has 
been triggered. The default is “controlling relay 1,” if 
relays are installed. Other selections are: 
controlling relay 2 (or 3 or 4 if installed) 
measurement #1 display flash  
measurement #2 display flash (or meas #3, #4, etc.., if 
measurements have been set up) 
Zero current output 1 (or 2 if installed) 
Note:  Additional menu items related to controlling relays are shown in following 
table. Other menu items may appear depending on entries you made 
elsewhere in the menus. 
Set up dist alarm 2 
(Alarm point, etc.)→→ 
 NEXT↓↓ 
After you set up an alarm, the option to set up the 
next alarm will be presented. If you don’t want to 
define another alarm, scroll past this screen. 