Operating Manual
Accu-Wave Chapter 4 Set up Additional Measurements 
TN Technologies 4-5 
0.000 cu ft 
 NEXT↓↓ 
material is computed based on the tank volume setup 
and the level measurement. The volume of the tank 
when full is required to compute the ullage (empty 
volume), the % full and the % empty values. 
“Set up Additional Measurements” Menu  
(Measurement Specific Parameters and Units) (cont.) 
Display  Comments 
If “mass” is selected, the following menu items are displayed. 
Mass  
units: Kgram 
 NEXT↓↓ CHANGE→→ 
Select the units for the mass measurement. Available 
units are: gram, kgram, pound, metric ton, short ton, 
long ton, and oz. 
material density 
0.000 Kgram/cu ft 
 NEXT↓↓ 
Enter the material density value. The density value 
(mass / volume) must be specified in the same units 
selected for mass and volume. Density values < 0.1 
are displayed in scientific notation and the units are 
not displayed due to display space constraints. 
If “rate” is selected the following menu items are displayed. 
rate readout will be 
rate of: 
Distance/time 
 NEXT↓↓ CHANGE→→ 
Select a (previously setup) measurement (distance in 
this example) for which to compute the rate of 
change. Measurement number for rate should be 
greater than the number of the base measurement. 
Smallest change for 
rate compute 
0.000 ft dist 
2.022E-2 assum NEXT↓↓ 
Rate will not be computed until the change in the 
measurement exceeds the value entered here. Once 
rate has been computed, it is recomputed when the 
change threshold is exceeded or when the expected 
time for the change threshold to be exceeded has 
elapsed. 
rate smoothing 
 factor 1.000 
{0.01=smoothest} 
{0.01 to 1.0} NEXT↓↓ 
Smoothing factor for rate measurement. A value of 
1.0 corresponds to no smoothing (current estimate is 
displayed), a value of 0.01 corresponds to the most 
smoothing (longest effective averaging time). 
rate time code 
ft dist/d 
 Continue↓↓ Change→→ 
Select the reference time interval for the rate 
measurement. Choices are: s (seconds), m (minutes), 
h (hours), d (days), w (weeks), M(months), and  
y (years). 
The next menu item is displayed only if “Service Only Items” are enabled. 
special code for 
measurement 2 0 
see manual 
 NEXT↓↓ 
Enter special measurement code. Typically used only 
for diagnostic purposes. 