Operating Manual
Chapter 4 Set up Additional Measurements  Accu-Wave 
4-8 TN Technologies 
you expect, the readout will still display the correct, scaled measurement value, as long as 
the scaled value can be displayed in four digits. In our example where 40,000 liters is 
scaled to read 40.00 on the display, if the gauge measures 43,875 liters, the displayed 
value would be 43.88. Any value up to 99,999 would be displayed correctly, rounded to 
four significant digits. 
Lowest Expected Reading 
This menu item is displayed if you set a highest expected reading value greater than 9999 
for a measurement. Use this item to enter the lowest expected measurement value (before 
display scaling). 
Scale Actual (highest expected reading) to High End Readout 
This menu item is displayed if you set a highest expected reading value greater than 9999 
for a measurement. Enter the desired readout value to be displayed for the highest expected 
reading. 
For example, to scale a highest expected reading value of 40,000 to a display value of 
40.00, enter 40.00 for the high-end readout value. 
Scale Actual (lowest expected reading) to Low End Readout 
This menu item is displayed if you set a highest expected reading value greater than 9999 
for a measurement. Enter the desired readout value to be displayed for the lowest expected 
reading. 
For example, to scale a lowest expected reading value of 30,000 to a display value of 
30.00, enter 30.00 for the low-end readout value. 
Set up custom units messages 
This menu item is displayed if you set a highest expected reading value greater than 9999 
for a measurement. By default, the original volume units will be displayed for the scaled 
value on the measurement display. The “Set up custom units messages” menu items allow 
you to set up a user-defined units message up to 10 characters long.  
In the above example, the volume measurement with units of liters was scaled by a factor 
of 1/1000 (40,000 is displayed as 40.00), so the displayed value has units of kiloliters. 
Thus, you might want to set up a custom units message to read “kiloliter.” Up to 8 custom 
messages can be defined using any combination of ASCII characters up to 10 characters in 
length. Refer to “Setting up Custom Units Messages” on page 8-20 for details. 
Position of Decimal 
Press the ← and → keys to select the position of the decimal point in the display readout. 
This is the last setting that is required for setting up the measurement. You can press ↓ 
followed by ← to return to the “additional measurements” menu, or press EXIT SETUP to 
save changes and return to the normal measurement display. 