Operating Manual
Chapter 4 Set up Additional Measurements  Accu-Wave 
4-10 TN Technologies 
Set up Alarm Menu 
Note: If relays are installed, the default indicator for alarm N is “controlling relay 
N,” even if N exceeds the number of relays. Press →→ to change the alarm 
indicator, and the “controlling relay #” options will then correctly reflect 
the number of installed relays. 
“Set up Alarm” Menu 
Display  Comments 
Set up alarm 3 
(Alarm point, etc.)→→ 
 NEXT↓↓ 
This item only shows up if an alarm (e.g., alarm 3) 
has been assigned to the measurement in the “Assign 
alarms to measurement” item of the “Set up fault 
alarms…” menu (see “Set Up Fault Alarms or 
Change Process Alarm” on page 6-3). The alarm 
number and the measurement number can be 
different. 
Note: The appearance of the following menu items depends on the selected 
measurement type. You must enter a set point to activate the rest of the Set 
up Alarm menu. Example displays show volume (cu ft) measurement. 
←←Exit alarm 3 setup 
Alarm 3 set point 
40.00 cu ft 
 NEXT↓↓ HELP→→ 
Enter the measurement value for the alarm set point; 
the value at which the alarm is activated.  
Alarm 3 clear based  
on clr point 
Chng to “dead band”→→ 
Continue as is.↓↓ 
Select a clear point or dead band configuration. A 
clear point is the exact measurement value at which 
the alarm is cleared. A dead band specifies the span 
between the set point and the implicit clear point. 
Alarm 3 clear point 
30.00 cu ft 
{Makes alarm “Hi” 
limit} NEXT↓↓ HELP→→ 
Enter desired clear point (or dead band) value, the 
measurement value at which the alarm is cleared 
(turned off). If dead band selected, enter the width of 
the dead band relative to the set point. 
Alm 3: cu ft  
is indicated by 
controlling relay 2 
 NEXT↓↓ CHANGE→→ 
Select action used to indicate that alarm 3 has been 
triggered. The default for alarm N is “controlling 
relay N,” if relays are installed. Other selections are: 
controlling relay 2 ( 3, 4 if installed) 
Meas #1 dispy flash (#2, #3, etc., if measurements 
have been setup) 
Zero current output 1 (or 2 if installed) 
Note:  Additional menu items related to controlling relays are shown in the next 
table. Other menu items may appear depending on entries you made 
elsewhere in the menus. 
Set up alarm 4 
(Alarm point, etc.)→→ 
 NEXT↓↓ 
If you assigned more than one alarm (e.g., alarm 4) to 
the measurement, menu items will be displayed 
allowing the option to set up these other alarms. If 
you don’t want to set up another alarm, scroll past 
this screen. 