Operating Manual
Chapter 5 Radar Fine Tuning Accu-Wave 
5-2 TN Technologies 
4. False Echo Management 
There can be several sources of radar echoes other than the material surface. If the 
automatic echo selection criteria are adequate for identifying the echo from the 
surface in your application, this menu provides tools to help the system select the 
proper echo.  
From the measurement display, use the → key to scroll to the “Radar fine tuning” menus. 
Press ↓ to enter this menu group and to scroll to the four menu groups listed above. Each 
of these four menu groups is discussed in more detail in the following sections. 
Material Type and Measurement Head Setup 
The “Material type and measurement head setup” menus allow you to enter information 
regarding the characteristics of your process material, the process seal, the system to 
determine the expected signal strength and any distance offset which might be caused by 
your specific process seal. Unless material information is entered, the system uses low 
initial amplification to prevent saturation. This configuration will serve in most 
applications due to the system’s very broad dynamic range. 
If the return signal is weak due to a combination of factors such as a low dielectric process 
material plus agitation, the situation may be improved by entering information about the 
material. If you select “is not” water based material, the system will increase the initial 
amplification to a moderate level. Entering the dielectric constant for your particular 
material will allow the system to optimize the initial amplification. 
If there are features in your tank that produce a stronger echo than the surface of the 
process material, you may see saturation warnings during operation. In this case, you will 
need to change the dielectric entry to a higher value. (The gain or amplification setting for 
the microwave signal is determined based on the value of the dielectric constant, as the 
value of the dielectric constant is increased, the amplification level is decreased.) Even 
though such offending signals may be stronger than the proper signal, the system can still 
be tuned to select the proper echo. It is important, however, that the initial amplification be 
set so that the strongest echo does not cause saturation. 
The “spectrum count to average” entry allows the system to enhance weak echoes by 
averaging several samples. An entry greater than 8 can slow the response to level changes. 
The spectrum averaging is performed prior to performing the echo limiting and echo 
rejection tests. Typically, this entry should be left at 1.0 (the default value). 
The “Material Type and Measurement Head Setup” menu items are summarized in the 
following table. 










