Operating Manual
Accu-Wave Chapter 5 Radar Fine Tuning 
TN Technologies 5-5 
“Modify Range of Measurement” Menu 
Display  Comments 
Distance REF LINE to 
highest possible 
level 0.000 ft 
 NEXT↓↓ HELP→→ 
Defines top of the ROI. If this entry is 0.0, the top of 
the ROI is set by the maximum level (or zero distance 
point) entry for the primary measurement. A non-zero 
entry here overrides the default value for the top of 
the ROI, and defines the closest echo processed. The 
gauge will not correctly measure above this level. 
Minimum value is 2.12 m (0.691 ft). 
REF LINE to maximum 
distance to be 
measured 0.000 ft 
 NEXT↓↓ HELP→→ 
Defines bottom of the ROI. If entry is 0.0, the bottom 
of the ROI is set by the zero level (or maximum 
distance point) entry for the primary measurement. A 
non-zero entry here overrides the default value for the 
bottom of the ROI, and defines the farthest echo 
processed. The gauge will not measure correctly 
below this level. Maximum value is 34 m (112 ft). 
Slope correction 
 factor 1.000 
 NEXT↓↓ HELP→→ 
Adjust gauge readout using following equations: 
 Distance = slope factor x (meas. value + offset). 
 Level = zero level point – distance 
Limit slope factor value to range of 0.995 and 1.005. 
Consult TN if this value needs to be changed. 
Direct Entry Code: ‘049003’ 
Distance offset  
correction  
0.000 ft 
 NEXT↓↓ HELP→→ 
Offset value for gauge readout adjustment equations 
(see entry above). Mounting imperfections can cause 
up to 0.10 ft offset error. Limit offset value to range 
of 0.0 to 0.16 ft (0.05m). Consult TN if this value 
needs to be changed. 
Direct Entry Code: ‘031003’ 