Operating Manual
Accu-Wave Chapter 5 Radar Fine Tuning 
TN Technologies 5-9 
“Response Time, Lost Echo and Agitator Reject Setup” Menu 
Display  Comments 
Level Signal 
time constant 
4 seconds 
 NEXT↓↓ HELP→→ 
Time constant for smoothing filter applied to the 
primary measurement (level or distance). Range of 
time constant value is from 1 to 1000 seconds. The 
longer the time constant, the greater the smoothing, 
and the slower the response of the system. Default 
value is 4 seconds. 
Direct Entry Code: ‘007004’ 
echo loss mode is 
hold last value 
 NEXT↓↓ CHANGE→→ 
When in “lost echo” mode (when gauge loses the 
return signal), display reports: “Echo lost.” Select one 
of the following actions: 
hold last value 
go to hold value (enter value in next menu item) 
go to maximum possible distance 
go to zero distance point 
Direct Entry Code: ‘142002’ 
Level hold mode 
& lost echo value 
0.000 ft level 
 NEXT↓↓ 
Only displayed when “go to hold value” is selected 
for the echo loss mode in the previous menu item  
Direct Entry Code: ‘037003’ 
time after echo  
lo
ss before going to  
lost echo mode  
4 seconds NEXT↓↓ 
Delay time between echo loss and switch to “lost 
echo” mode. During this period display will report: 
“Waiting for echo.” Enter the length of time the 
gauge waits before going to the signal loss mode 
selected above. The default is 4 seconds. 
Direct Entry Code: ‘029004’ 
Agitator reject time  
1.000 sec 
(range gate delay) 
 NEXT↓↓ HELP→→ 
Agitator reject time is the delay time before accepting 
an echo outside the range gate. The range gate is a 
window that moves with the “measured” level. 
During this period display reports: “Waiting good 
echo.” Used to reject false echoes caused by an 
agitator blade passing through the signal. 
Direct Entry Code: ‘044003’ 
height of range gate 
(moving window) 
0.000 ft 
 NEXT↓↓ HELP→→ 
Enter height of range gate which is centered on the 
distance measurement readout. A larger window will 
admit a larger number of signals as possibly valid. If 
this variable is set to 0.0, the gauge will use the 
default value of 25cm (± 12.5cm) range window. 
Direct Entry Code: ‘019003’ 