Operating Manual
Chapter 5 Radar Fine Tuning Accu-Wave 
5-18 TN Technologies 
by taking new clutter data, selecting “Do” disable clutter erase, or entering a clutter 
distance of zero (which disables the clutter erase feature). 
The need for clutter erase can be established by viewing the echo report. Look for echoes 
within a few feet (up to 7 ft or 2 m). If the signal strength associated with these echoes is 
more than 5% of the signal strength from the material surface, you should consider doing 
clutter erase. If the signal strength associated with these echoes is more than 50% of the 
signal strength from the material surface, you should verify the sensor mounting. 
The clutter distance may be entered or changed before or after clutter data is taken. 
However, the clutter distance must always be at least 30 cm (1 ft) less than the distance to 
the material surface when clutter data was taken. In addition, the clutter distance must not 
be extended to include any true physical sources of echoes (such as pipes) that might be 
covered during normal operations. 
Automatic Clutter Update 
In most applications which require clutter erase, the clutter data is only taken once. 
However, in applications where the clutter echoes are strong and the clutter can change 
due to buildup or ambient temperature extremes, automatic timed clutter data may be 
employed. 
Timed clutter data is activated by entering a clutter update time from a minimum of 1 
minute to a maximum value of 45 days (65000 minutes). At the end of each clutter update 
interval, the system activates the clutter update process. The system waits to measure the 
clutter data until the measured level is at least 30 cm (1 ft) beyond the value entered for the 
clutter distance. 