Accu-Wave Chapter 5 Radar Fine Tuning 
TN Technologies 5-27 
“Dynamic Threshold Setup” Menu 
[“Service Only Items” must be enabled] 
Display  Comments 
Do not enable 
dynamic threshold 
Change to “Do”→→ 
Continue as is.↓↓ 
When set to default value of “Do not enable dynamic 
threshold,” the menu items below are not displayed. 
Change to “Do” to enable dynamic threshold. Reset to 
“Do not” at any time to disable dynamic threshold. 
Direct Entry Code: ‘026060’ 
Change 
dynamic threshold 
by .8000 each sec 
 NEXT↓↓ 
If there is no acceptable echo with an amplitude 
greater than the present value of the dynamic 
threshold, the dynamic threshold is decreased by the 
factor entered here (value must be less than 1.0). 
Direct Entry Code: ‘010003’ 
dynamic threshold 
base= signal str/2^1 
 NEXT↓↓ 
The initial baseline value of the dynamic threshold is 
set at the amplitude of the previously accepted echo 
divided by 2
K
. Enter the value of the exponent K. For 
example, if 3 is entered, the initial baseline value will 
be 1/8 (1/2
3
) of the amplitude of the previously 
accepted echo. 
Direct Entry Code: ‘058004’ 
Dynamic threshold 
increase smoothing 
factor 1.000 
{0.001 to 1.0} NEXT↓↓ 
Enter value for the “increase smoothing factor” (ISF). 
Default value is 1.0 or no smoothing. If the new base 
threshold value is greater than the old threshold, the 
base value of the dynamic threshold will be equal to:  
 threshold
old 
+ [ISF x (threshold
new
 – threshold
old
)] 
Direct Entry Code: ‘149013’ 
Dynamic threshold  
decrease smoothing 
factor 1.000 
{0.001 to 1.0} NEXT↓↓ 
Enter value for the “decrease smoothing factor” 
(DSF). Default value is 0.1. If the new base threshold 
value is less than the old threshold, the base value of 
the dynamic threshold will be equal to:  
 [threshold
old 
+ [DSF x (threshold
new
 – threshold
old
)] 
Direct Entry Code: ‘149023’ 
Overall echo selection process 
Note: The following example illustrates how the Radar Fine Tuning tools work in 
combination with one another. It is not intended to reflect the setup of a 
typical application. 
Range of Interest (ROI) Setup: 
• Primary Measurement:  Level 
• Distance REF LINE to zero level point:  25.00 ft 