Operating Manual
Chapter 6 Cuurent Outputs, Relays, and Alarms Accu-Wave 
6-2 TN Technologies 
“Modify or Re-assign Current Output” Menu 
Display  Comments 
Modify or re-assign 
current output↓↓ 
 Other functions→→ 
Press ↓ to access “Current Output” menu items. Press 
→ to scroll to other menus. 
maximum current 
output 20.00 mA 
(4.000 to 20.00) 
 NEXT↓↓ 
The range for the maximum current output is from 
the minimum current output (currently 4.000mA) to 
20.00mA. The default value for the maximum current 
output is 20.00mA. 
minimum current 
output 4.00 mA 
(.0001 to 20.00) 
 NEXT↓↓ 
The range for the minimum current output is from 
.0001mA to the maximum current output (currently 
20.00mA). The default value for the minimum 
current output is 4.000 mA. 
Note:  Entering a value of exactly 0.0 causes the 
default value of 4.000mA to be used for the 
minimum current output. 
Note: If two or more measurements have been set up, the next two menu items 
will be displayed. If a second current output is also installed, the next two 
menu items will be repeated for “current out 2.” 
Mea 1: ft level 
is sent to current 
out 1 in normal 
mode. NEXT↓↓ CHANGE→→ 
Assign desired measurement to current output 1 in 
the normal mode. Select from the primary 
measurement and the measurements set up using the 
“Additional measurements” menus. 
Mea 1: ft level 
is sent to current 
out 1 in alternate 
mode. NEXT↓↓ CHANGE→→ 
Assign a measurement to current output 1 in the 
alternate mode. If system includes more than one 
current output, additional entries will show up. 
Note: If a second current output is installed, the next two menu items will be 
displayed first for “current output 1” and then for “current output 2.” 
correction factor 
for current output 
at maximum: 1.000 
 NEXT→→ 
Fine tune the maximum current output value to 
correct for component and wiring variations. (The 
maximum current output value is scaled by this 
value.) This item is repeated for each current output 
installed. 
correction factor 
for current output 
at minimum: 1.000 
 NEXT→→ 
Fine tune the minimum current output value to 
correct for component and wiring variations. (The 
minimum current output value is scaled by this 
value.) This item is repeated for each current output 
installed. 
current output 
hold mode value 
50.00 % of scale 
 NEXT↓↓ 
Enter desired value for mid-range hold value for the 
current output. Value is entered as a percentage of the 
maximum current output value. The default is 50%. 










