Operating Manual
Chapter 7 Action Items Accu-Wave 
7-4 TN Technologies 
Hold Action Items 
“Hold Action Items” Menu 
Display  Comments 
“Hold” action items 
(Hold reading, etc)→→ 
 NEXT↓↓ 
Press → to enter the “Hold action items” menu. Press 
↓ to continue to next “Action Item” menu subgroup. 
Clear all holds 
(none now in effect) 
←←Exit this menu. 
 NEXT↓↓ EXECUTE CMD→→ 
Execute this command to clear all holds (if any holds 
are in effect). 
Hold current output 
at max (20.00 mA) 
←←Exit this menu. 
 NEXT↓↓ EXECUTE CMD→→ 
Holds all current outputs at the value entered for the 
maximum current output. 
Hold current output 
at min (4.000 mA) 
←←Exit this menu. 
 NEXT↓↓ EXECUTE CMD→→ 
Holds all current outputs at the value entered for the 
minimum current output. 
Hold current output 
at zero mA 
←←Exit this menu. 
 NEXT↓↓ EXECUTE CMD→→ 
Holds all current outputs at 0.0 mA. 
Current output 
hold mode value 
50.00% of scale 
 NEXT↓↓ 
Enter value in % for the mid-range current output 
hold value. Default value is 50.00%. 
Hold current output 
at 50.00% of scale 
←←Exit this menu. 
 NEXT↓↓ EXECUTE CMD→→ 
Press → to execute command to hold current output at 
the mid-range value specified in the previous menu 
item. 
Level hold mode 
 value 
0.000 ft level 
 NEXT↓↓ 
Enter the hold value for the primary measurement 
(level or distance). This value is also used as the “lost 
echo value” when “go to hold value” is selected as the 
“echo loss mode.” See page 5-8. 
Hold level at 
0.000 ft level 
←←Exit this menu. 
 NEXT↓↓ EXECUTE CMD→→ 
Press → to hold primary measurement value at the 
hold value specified in the previous menu item. After 
executing command, display changes to read “Clear 
holds→.” Press → to clear the hold level command. 