Instruction Manual

Operating Instructions
for
CaterTemp Plus &
Therma Plus
1762 W. 20 S. #100, Lindon, UT 84042
Phone: (801) 756-7705 Fax: (801) 756-8948
info@thermoworks.com
www.thermoworks.com
Instrument Record Card
Model Order # Range Resolution
CaterTemp Plus 221-056 -50.0ºC to +299.9ºC 0.1ºC/ºF
-58.0ºF to +575.0ºF
CaterTemp Plus F/R 221-059 -50.0ºC to +299.9ºC 0.1ºC/ºF
-58.0ºF to +575.0ºF
Therma Plus 221-071 -100.0ºC to +1372ºC 0.1/1ºC/ºF
-148.0ºF to +2502ºF
Calibration
Test Temperature
0°C/32°F 100°C/212°F 1000°C/1832°F
Instrument Indication
Serial Number
Calibrated by Date
This instrument has been checked or calibrated against reference instrument(s) calibrated
by a UKAS Accredited Calibration Laboratory which are traceable via International
Agreement to all major National Standards including NPL and NIST.
To BS EN 60584 : 1996
Manufactured in the UK by E.T.I. Ltd. for
CaterTemp Plus: Therma Plus:
Temp Range: -58° to 575°F
(-50° to 299.9°C)
-148°F to 2502°F
(-100°C to 1372°C)
Accuracy: Meter & Probe
±0.5°F (±0.28°C) from 32°F to
212°F (0°C to 100°C)
±0.72°F (±0.4°C), ±0.1% full
range
Meter Only
±0.5°F (±0.28°C) from 32°F to
212°F (0°C to 100°C)
±0.72°F (±0.4°C), ±0.1% full
range
Resolution: 0.1°F or °C full range 0.1°F or °C auto ranging at
599.9°F or 299.9°C
Display Units: °F or °C switchable
Ambient Range: -4°F to 122°F (-20°C to 50°C)
Probe Type: Type K Thermocouple (1/8” diam
x 4”L) reduced tip
Type K Thermocouple
Cold Junction
Stability:
±0.025°C/°C maximum
Response Time: 3-4 seconds to full reading
including probe
1 seconds maximum update
Housing: Extruded aluminum (IP66/67 rated)
Power: 9V Battery (included)
Battery Life: 5,000 hours
Over Range
Indication:
‘HI’
Open Circuit
(probe damage):
‘Err’
Dimensions: 1.4” x 2.4” x 4.5” (35mm x 60mm x 115mm)
Weight: 0.5 lbs (244g)
Warranty: 2-years for meter,
3-months for probe
2-years for meter
Calibration: NIST-traceable certificate included