12. Calibrating the Microscope
Focus the microscope on the calibration ruler.
Rotatethenefocusdial(A6) until the picture becomes sharp.
Adjust the microscope position so the image is lined
up square with the XY guides.
Double-click an image in the saved images window (B7) to enter the measurement interface, then select
the calibration icon.
Click and drag to select start and end points across
a known value e.g. 1 mm increment along the ruler.
When the dialog box shows, enter the exact value of
the measurement sample in the Actual dimension
eld.Thesoftwarewillcalibratethemagnication.
Select OK to conrm. The adjusted magnication
value will show in the upper right corner.
13. Care and Cleaning
Disconnect the product from your computer before each cleaning. If it is hot, wait for it to cool down.
Do not use any aggressive cleaning agents, rubbing alcohol or other chemical solutions as they can cause damage
to the housing and malfunctioning.
The LED lights are not replaceable.
Avoid touching the lens with any substance. Clean the lens periodically with low pressure compressed air.
Cleantheproductwithadry,bre-freecloth.