P200 Manual V1.0

@ToolkiRC 2021
14
other functions
1. Firmware upgrade
After connecting the P200 to the computer with the
USB data cable in the box, the computer will recognize a
U disk named Toolkit. Download the upgrade file
app.upg on the official website to overwrite the files in
the U disk to upgrade the firmware.
2. Fan level
When the internal temperature of the device exceeds
45°C, the fan turns on half-speed air volume to reduce
noise. When the internal temperature exceeds 53°C, the
fan turns on full-speed air volume to enhance heat
dissipation.
3. Manually calibrate voltage and current
Press the power button, within 1 second,
simultaneously press and hold [A dial] + [V dial] without
releasing, the system will enter the manual voltage
calibration function. Use a voltmeter to measure the
actual input and output voltage, move the cursor to the
corresponding voltage value, modify the voltage value
to be consistent with the voltmeter value, and achieve
calibration. After the calibration is completed, move the
cursor to save, short press once, the buzzer will beep for
a long time, and the save is successful. Just exit or shut
down.