User guide

11
ENGLISH
TOPCOM BPM Arm 5500 XLD
10.1 Time adjustment
11 Troubleshoot
To adjust the date an time:
Press the start/stop and memory button
simultaneously. The month will blink on the display.
Press the memory button successively to change
the month.
Press the start/stop button to confirm the month.
The day will blink on the display.
Change the day, hours and minutes as described above,
using the memory button to change and the start/
stop button to confirm every setting.
After setting the minutes, the device is ready to use.
Error message
on display
Reason Solution
Batteries low Change the batteries
Er0 Pressure system is unstable before
measurement.
Don't move and try again.
Er1 Systolic pressure detect fail
Er2 Diastolic pressure detect fail
Er3 Pneumatic system blocked or cuff is
too tight during inflation
Place the cuff correctly and
try again.
Er4 Pneumatic system leakage or cuff is
too loose during inflation
Er5 Inflation above 300mmHg
Measure again after 5
minutes. If the monitor is still
abnormal, please contact the
local distributor or the
factory.
Er6 Over 3 minutes while pressure
above 15 mmHg
Er7 EEPROM accessing error
Er8 Device parameter check error
Er9 MCU self-verity error
ErA Span error: pressure sensor
parameter error
DIA
mmHg
SYS
mmHg
PUL/min