User guide

- 12 -
Never enable the “Batter” (battery) function when operating the scale on conventional
disposable AA Batteries. The “Battery” function may be enabled only when using rechargeable
multi-use batteries.
Useful Tips for Operating the Scale with a Battery
In order to increase the battery’s operating life, it is recommended to set the scale’s back light mode to
(“back light battery” – B_LBAT). Refer to Chapter 28 (LCD back light).
When operating the unit on batteries it is recommended to enable the scale’s “Auto-Off” function by
setting it to AUT BAT. This will turn the scale off automatically after 5 minutes of inactivity. Refer to
Chapter 29 (Auto Off).
To install or replace the scale’s batteries, follow the steps below:
1. Remove the scale’s pan, pan base, as
well as the draft ring (BTA210
Models) and carefully place the scale
up-side down.
2. Remove the security screws (11) and
lift the battery compartment cover
(12).
3. Remove the battery nest from the
compartment (13).
4. Install four AA batteries. Be sure
position the batteries with correct
polarity.
Chapter 7: Battery Installation, Replacement, and Operation