User Manual

- 47 -
This function is used when vibration is encountered during weighing. This may be the result of ground
vibration, or a vibrating live load on the pan. Increasing the filter setting will increase the time required for a
measurement. The integration time of the measurement is increased thereby eliminating higher frequency
vibration.
1. Press the MENU key, when the option “Filter” is displayed press the T key.
2. The scale will display options “Off” and available filter sensitivity levels “Fil 10, Fil 20, Fil 30 and Fil
40”. To select and enable Filtering press the T key when a desired filter level is displayed.
Filter Sensitivity
Fil 10 Low
Fil 20 Medium
Fil 30 High
Fil 40 Severe
3. To disable filtering select “Fil Off” from the filter menu.
Cha
p
ter 21: Filterin
g
(
FILter
)
fILTER
FIL 10 FIL 20 FIL 30 FIL 40