9.2.2 50 % Flow Rate Calibration
Using Mag-Provers internal calibration circuit, 50% flow rate calibration can be performed. For
calibration procedure, see the calibration procedure for 0% flow rate. (For 50% flow rate calibration,
select "CAL 50" from the setting menu.)
9.2.3 100 % Flow Rate Calibration (Span Calibration)
Using Mag-Prover’s internal calibration circuit, 100% flow rate calibration can be performed. For
calibration procedure, see the calibration procedure for 0% flow rate. (For 100% flow rate calibration,
select "CAL 100%" from the setting menu.)
9.2.4 Checking the Excitation Current
You can monitor the exciting current value.
Checking the exciting current value
Switch operation
Display example
Description
R:CAL
1 CAL 0%
2 CAL 50%
3 CAL 100%
4 EXC DSPL
5 EXIT
Select "EX DSPL" in the setting item
selection screen.
STEP1
EXC DSPL
0.1998
A
The excitation current value appears.
Pushing returns you to the setting
menu.
ESC
ESC
ESC